5.9.2 View Curve
87
|
www.lakeshore.com
5.9.2 View Curve
The View Curve operation provides read-only access to all standard and user curves.
To perform the View Curve operation follow this procedure.
1. Press
Curve Entry
, scroll to View Curve, then press
Enter
.
2. Scroll to the desired curve and press
Enter
again to view the curve header infor-
mation.
3. To view the curve breakpoints, highlight the Curve Points parameter and press
Enter
. The list of breakpoint pairs is scrollable, but data cannot be edited.
4. Press
Escape
(Exit Menu) to return to the curve header parameter list.
5. Press
Escape
(Exit Menu) again to exit the Curve Entry menu and return to nor-
mal operation.
Menu Navigation:
Curve Entry
Q
View Curve
Interface Command:
CRVHDR, CRVPT
5.9.3 Erase Curve
You can erase user curves that are no longer needed. Erase Curve sets all identifica-
tion parameters to default and blanks all breakpoint values.
To perform the Erase Curve operation follow this procedure.
1. Press
Curve Entry
, scroll to Erase Curve, then press
Enter
.
2. Scroll to the desired curve and press
Enter
.
3. Choose Yes at the confirmation message to finalize the operation.
4. To cancel the operation, either choose No to the confirmation message, or press
Escape
.
Menu Navigation:
Curve Entry
Q
Erase Curve
Q
(
21–59
)
Interface Command:
CRDEL
5.9.4 Copy Curve
Temperature curves can be copied from one location inside the Model 336 to another.
This is a good way to make small changes to an existing curve. Curve copy may also be
necessary if you need the same curve with two different temperature limits or if you
need to extend the range of a standard curve. The curve that is copied from is always
preserved.
The copy routine allows you to overwrite an existing user curve. Please ensure the curve
number you are writing to is correct before proceeding with the copy curve operation.
1. To perform the Copy Curve operation press
Curve Entry
, scroll to Copy Curve, then
press
Enter
.
2. Scroll to the desired curve to copy, and press
Enter
. A list of user curves is dis-
played.
3. Scroll to the desired user curve location to copy to, and press
Enter
.
4. Choose Yes at the confirmation message to finalize the operation.
5. To cancel the operation, either choose No to the confirmation message, or press
Escape
.
Menu Navigation:
Curve Entry
Q
Copy Curve
Q
(
1–59
)
Q
(
21–59
)
Interface Command: (No interface command directly corresponds to the copy curve
operation. You can use the CRVHDR and CRVPT commands to read curve information
from one curve location and write that information to another curve location.)
Содержание 336
Страница 4: ...Model 336 Temperature Controller...
Страница 6: ...Model 336 Temperature Controller...
Страница 26: ...14 cHAPTER 1 Introduction Model 336 Temperature Controller...
Страница 54: ...42 cHAPTER 3 Installation Model 336 Temperature Controller...
Страница 84: ...72 cHAPTER 4 Operation Model 336 Temperature Controller...
Страница 104: ...92 cHAPTER 5 Advanced Operation Model 336 Temperature Controller...
Страница 164: ...152 cHAPTER 7 Options and Accessories Model 336 Temperature Controller...
Страница 178: ...166 cHAPTER 8 Service Model 336 Temperature Controller...