1.3 Model 336 Specifications
7
|
www.lakeshore.com
1.3 Model 336
Specifications
1.3.1 Input
Specifications
Standard
inputs and
scanner option
Model 3062
Sensor
Tempera-
ture Coeffi-
cient
Input Range
Excitation
Current
Display
Resolution
Measurement
Resolution
Electronic
Accuracy
(at 25 °C)
Measurement Temperature
Coefficient
Electronic
Control Stability
8
Diode
Negative
0 V to 2.5 V
10 µA ±0.05%
9,10
100 µV
10 µV
±80 µV ±0.005%
of rdg
(10 µV + 0.0005% of rdg)/°C
±20 µV
Negative
0 V to 10 V
10 µA ±0.05%
9,10
100 µV
20 µV
±320 µV ±0.01%
of rdg
(20 µV + 0.0005% of rdg)/°C
±40 µV
PTC RTD
Positive
0
)
to 10
)
1 mA
11
0.1 m
)
0.2 m
)
±0.002
)
±0.01% of rdg
(0.01 m
)
+ 0.001% of rdg)/°C
±0.4 m
)
0
)
to 30
)
1 mA
11
0.1 m
)
0.2 m
)
±0.002
)
±0.01% of rdg
(0.03 m
)
+ 0.001% of rdg)/°C
±0.4 m
)
0
)
to 100
)
1 mA
11
1 m
)
2 m
)
±0.004
)
±0.01% of rdg
(0.1 m
)
+ 0.001% of rdg)/°C
±4 m
)
0
)
to 300
)
1 mA
11
1 m
)
2 m
)
±0.004
)
±0.01% of rdg
(0.3 m
)
+ 0.001% of rdg)/°C
±4 m
)
0
)
to 1 k
)
1 mA
11
10 m
)
20 m
)
±0.04
)
±0.02% of rdg
(1 m
)
+ 0.001% of rdg)/°C
±40 m
)
0
)
to 3 k
)
1 mA
11
10 m
)
20 m
)
±0.04
)
±0.02% of rdg
(3 m
)
+ 0.001% of rdg)/°C
±40 m
)
0
)
to 10 k
)
1 mA
11
100 m
)
200 m
)
±0.4
)
±0.02% of rdg
(10 m
)
+ 0.001% of rdg)/°C
±400 m
)
NTC RTD
10 mV
Negative
0
)
to 10
)
1 mA
11
0.1 m
)
0.2 m
)
±0.002
)
±0.06%
of rdg
(0.01 m
)
+ 0.001% of rdg)/°C
±0.3 m
)
0
)
to 30
)
300 µA
11
0.1 m
)
0.2 m
)
±0.002
)
±0.06%
of rdg
(0.03 m
)
+ 0.001% of rdg)/°C
±0.9 m
)
0
)
to 100
)
100 µA
11
1 m
)
1 m
)
±0.01
)
±0.04%
of rdg
(0.1 m
)
+ 0.001% of rdg)/°C
±3 m
)
0
)
to 300
)
30µA
11
1 m
)
2 m
)
±0.01
)
±0.04% of
rdg
(0.3 m
)
+ 0.001% of rdg)/°C
±9 m
)
0
)
to 1 k
)
10 µA
11
10 m
)
10 m
)
+0.002%
of rdg
±0.1
)
±0.04%
of rdg
(1 m
)
+ 0.001% of rdg)/°C
±30 m
)
±0.004%
of rdg
0
)
to 3 k
)
3 µA
11
10 m
)
20 m
)
+0.002%
of rdg
±0.1
)
±0.04%
of rdg
(3 m
)
+ 0.001% of rdg)/°C
±90 m
)
±0.004%
of rdg
0
)
to 10 k
)
1 µA
11
100 m
)
100 m
)
+0.002%
of rdg
±1.0
)
±0.04%
of rdg
(10 m
)
+ 0.001% of rdg)/°C
±300 m
)
±0.004% of rdg
0
)
to 30 k
)
300 nA
11
100 m
)
200 m
)
+0.002%
of rdg
±2.0
)
±0.04%
of rdg
(30 m
)
+ 0.001% of rdg)/°C
±900 m
)
±0.004% of rdg
0
)
to 100 k
)
100 nA
11
1
)
1
)
+0.005% of rdg ±10.0
)
±0.04%
of rdg
(100 m
)
+ 0.001% of rdg)/°C
±3
)
±0.01% of
rdg
8
Control stability of the electronics only, in ideal thermal system
9
Current source error has negligible effect on measurement accuracy
10
Diode input excitation can be set to 1 mA
11
Current source error is removed during calibration
12
Accuracy specification does not include errors from room temperature compensation
TABLE 1-3
Input specifications
Содержание 336
Страница 4: ...Model 336 Temperature Controller...
Страница 6: ...Model 336 Temperature Controller...
Страница 26: ...14 cHAPTER 1 Introduction Model 336 Temperature Controller...
Страница 54: ...42 cHAPTER 3 Installation Model 336 Temperature Controller...
Страница 84: ...72 cHAPTER 4 Operation Model 336 Temperature Controller...
Страница 104: ...92 cHAPTER 5 Advanced Operation Model 336 Temperature Controller...
Страница 164: ...152 cHAPTER 7 Options and Accessories Model 336 Temperature Controller...
Страница 178: ...166 cHAPTER 8 Service Model 336 Temperature Controller...