Lake Shore Model 420 Gaussmeter User’s Manual
Introduction
1-1
CHAPTER 1
INTRODUCTION
1.0 GENERAL
Lake Shore Cryotronics, Inc. designed and manufactures the Model 420 in the United States of America. It is
highly accurate and ideally suited for field work. It features:
• Field Measurement:
– DC Accuracy: ±0.45% of Reading ±0.05% of Range.
– Auto
Range.
– DC or AC Field Measurement.
– Individual Linearization of Hall Probes.
• Alphanumeric Display:
– 4-character, 1 Part In 3,000 Resolution On All Ranges.
– 2 Line by 16 Character Backlighted Liquid Crystal Display.
• Other Operating Functions:
– Gauss or Tesla Units.
– Max
Hold.
– Relative
Reading.
• Interface:
– Serial Interface (RS-232C Electrical Format).
– Corrected and Monitor Analog Outputs.
• Probe Compatibility:
– High Stability Probes (HST) – 300 G to 300 kG Full-Scale Ranges.
– High Sensitivity Probes (HSE) – 30 G to 30 kG Full-Scale Ranges.
– Ultra High Sensitivity Probes (UHS) – 300 mG to 30 G Full-Scale
Ranges.
We welcome comments concerning this manual. Although we make every effort to keep it error-free, some
may occur. When reporting a specific problem, describe it briefly and include the appropriate paragraph,
figure, table, and page number. Send comments to Lake Shore Cryotronics, Attn: Technical Publications, 575
McCorkle Blvd., Westerville, Ohio 43082-8888. This manual is subject to change without notice.
1.1 MODEL 420 GAUSSMETER SYSTEM DESCRIPTION
The Model 420 has an easy-to-use front panel that includes an alphanumeric Liquid Crystal Display (LCD) for
message-based front panel operation. Perform and monitor most operations through the front panel keypad
and message display. The display has a resolution of 4-characters.
The Model 420 measures fields from 3 mG (0.3 µT) to 300 kG (30 T), in either gauss or tesla. Set field ranges
either manually or with auto ranging. The
Range
key displays the full scale value and cycles through
allowable ranges for the probe installed. The Auto Range function automatically selects the range with the
best resolution for the measured field.
The gaussmeter measures both DC and AC field values. In DC operation, it displays the field at the probe
with the sign (orientation), followed by appropriate field units. In AC operation, it displays a RMS value for the
field at the probe. The DC and AC values are available over the Serial I/O Computer Interfaces and also the
Corrected Analog Output. The Monitor Analog Output displays actual magnetic field waveform.
The Max Hold function captures and displays the largest field magnitude since the last Max Reset. After
pressing
Max Hold
, the maximum value appears in the lower display while the upper display contains the live
field reading. In AC RMS, the Max Hold feature displays the maximum RMS value of the waveform.
The relative function displays small variations in larger fields. The
Relative Set
key captures or defines a
relative setpoint which becomes the zero point of the relative reading and appears on the lower line of the
display. The relative reading appears in the top display with a
▲
symbol. The relative reading maintains the
same resolution and units as the setpoint.
Accept new settings in two ways: 1) Wait a few seconds until the built in timeout returns you to the normal
display, or 2) Press any key other than the function currently addressed. For example, after changing the
range, press any key other than Range.
Содержание 420
Страница 6: ...Lake Shore Model 420 Gaussmeter User s Manual iv Table of Contents This Page Intentionally Left Blank ...
Страница 14: ...Lake Shore Model 420 Gaussmeter User s Manual 2 4 Installation This Page Intentionally Left Blank ...
Страница 22: ...Lake Shore Model 420 Gaussmeter User s Manual 3 8 Operation This Page Intentionally Left Blank ...
Страница 31: ...Lake Shore Model 420 Gaussmeter User s Manual Remote Operation 4 9 This Page Intentionally Left Blank ...
Страница 32: ......
Страница 46: ...Lake Shore Model 420 Gaussmeter User s Manual A 2 Appendix A This Page Intentionally Left Blank ...