![Lake Shore 420 Скачать руководство пользователя страница 36](http://html.mh-extra.com/html/lake-shore/420/420_user-manual_670633036.webp)
Lake Shore Model 420 Gaussmeter User’s Manual
5-4
Service
5.5 OPERATING SOFTWARE EPROM REPLACEMENT
The operating software for the Model 420 is contained on one Erasable Programmable Read Only Memory
(EPROM) Integrated Circuit (IC). The EPROM is numbered U36 and located just behind the microprocessor
IC (U31). The EPROM also has a label on top identifying the software version and date (see Figure 5-9). Use
the procedure below to replace the operating software EPROM.
WARNING: To avoid potentially lethal shocks, turn off the instrument and disconnect it from the AC
power line before performing this procedure.
1.
Set power switch to Off and disconnect power cord from rear of unit.
2.
Use 5/64 hex key to remove four screws attaching top panel to unit.
3.
Use 5/64 hex key to loosen two screws attaching bottom panel to unit.
4.
Carefully remove back bezel.
5.
Slide top panel back and remove from unit.
6.
Locate software EPROM U36 on main circuit board. Note its orientation (circular notch on front of IC).
See Figure 5-9.
7.
Use IC puller to remove existing EPROM from socket.
8.
Use IC insertion tool to place new EPROM into socket, noting its orientation.
9.
Replace top of enclosure and secure with four screws.
10.
Replace back bezel and use 5/64 hex key to tighten two screws attaching bottom panel to unit.
11.
Reconnect power cord to rear of unit and set power switch to On.
12.
Perform the initial setup and system checkout (see Paragraph 2.6). When replacing the operating
software EPROM, all operating parameters reset to factory defaults specified in Paragraph 3.14.
Front
R
ear
Transformer
Fuse
Power Inlet
U36
Operating Software
EPROM
C-450-U-5-9
Figure 5-9 Location Of Operating Software EPROM
Содержание 420
Страница 6: ...Lake Shore Model 420 Gaussmeter User s Manual iv Table of Contents This Page Intentionally Left Blank ...
Страница 14: ...Lake Shore Model 420 Gaussmeter User s Manual 2 4 Installation This Page Intentionally Left Blank ...
Страница 22: ...Lake Shore Model 420 Gaussmeter User s Manual 3 8 Operation This Page Intentionally Left Blank ...
Страница 31: ...Lake Shore Model 420 Gaussmeter User s Manual Remote Operation 4 9 This Page Intentionally Left Blank ...
Страница 32: ......
Страница 46: ...Lake Shore Model 420 Gaussmeter User s Manual A 2 Appendix A This Page Intentionally Left Blank ...