Appendix A
A-7
Active Area
0.175 in.
(max.)
0.09 in. (max.)
+B
1.25 in.
5 ±0.5 in.
(Lead Length)
+
Figure B-7 Axial Hall Generator HGA-2010 Dimensions
Table B-2 Axial Hall Generator Specifications
Axial HGA-2010 HGA-3010 HGA-3030
Description
General purpose axial; high
sensitivity
Instrumentation quality axial;
low temperature coefficient;
phenolic package
Instrumentation quality axial;
phenolic package
Active area
0.005 x 0.005 inch square
0.030 inch diameter circle
0.030 inch diameter circle
Input resistance
450 to 900 ohms
1 ohm
2 ohms
Output resistance
550 to 1350 ohms
1 ohm
2 ohms
Nominal control current
(I
CN
)
1 mA
100 mA
100 mA
Maximum continuous
current (non-heat sinked)
10 mA
300 mA
300 mA
Magnetic sensitivity (I
C
=
nominal control current)
11 to 28 mV/kG
0.56 to 1.04 mV/kG
6.0 to 10.0 mV/kG
Maximum linearity error
(sensitivity versus field)
±1% RDG (-10 to +10 kG)
±2% RDG (-20 to +20 kG)
±1% RDG (-20 to +20 kG)
±1.5%
RDG
(-100
to
+100
kG)
±0.25% RDG (-10 to +10 kG)
±1.00% RDG (-30 to +30 kG)
Zero field offset voltage
(I
C
= nominal control
current)
±2.8 mV (max.)
±50 µV (max.)
±75 µV (max.)
Operating temperature
range
–20 to 75 °C
–40 to +100 °C
–40 to +100 °C
Mean temperature
coefficient of magnetic
sensitivity
–0.06%/°C (max.)
±0.005%/°C (max.)
–0.04%/°C (max.)
Mean temperature
coefficient of offset (I
C
=
nominal control current)
±1 µV/°C (max.)
±0.4 µV/°C (max.)
±0.3 µV/°C (max.)
Mean temperature
coefficient of resistance
+0.15%/°C (approx.)
±0.15%/°C (approx.)
+0.18%/°C (approx.)
Leads
34 AWG copper with poly-
nylon insulation
34 AWG copper with poly-
nylon insulation
34 AWG copper with poly-
nylon insulation
Содержание 420
Страница 6: ...Lake Shore Model 420 Gaussmeter User s Manual iv Table of Contents This Page Intentionally Left Blank ...
Страница 14: ...Lake Shore Model 420 Gaussmeter User s Manual 2 4 Installation This Page Intentionally Left Blank ...
Страница 22: ...Lake Shore Model 420 Gaussmeter User s Manual 3 8 Operation This Page Intentionally Left Blank ...
Страница 31: ...Lake Shore Model 420 Gaussmeter User s Manual Remote Operation 4 9 This Page Intentionally Left Blank ...
Страница 32: ......
Страница 46: ...Lake Shore Model 420 Gaussmeter User s Manual A 2 Appendix A This Page Intentionally Left Blank ...