![Lake Shore 420 Скачать руководство пользователя страница 50](http://html.mh-extra.com/html/lake-shore/420/420_user-manual_670633050.webp)
Lake Shore Model 420 Gaussmeter User’s Manual
A-6
Appendix A
B5.0 SPECIFICATIONS
This section covers three types of Hall generators available from Lake Shore: HGCA & HGCT Series
Cryogenic Hall generators (Figures B-5 and B-6) with specifications (Table B-1), HGA Series Axial Hall
generators (Figures B-5 and B-7) with specifications (Table B-2), and HGT Series Transverse Hall generators
(Figures B-8 thru B-10) with specifications (Table B-3).
0.105 in.
+B
9.5 in.
(min.)
0.20 in.
0.25 in.
diameter
0.20 in.
diameter
Figure B-5 Axial Hall Generator HGA-3010, HGA-3030, and HGCA-3010 Dimensions
0.240 in.
(max.)
+B
0.63 in.
9.5 in (min.)
(Lead Length)
Protective
Ceramic Case
Center of
Active Area
0.180 in.
0.043 in. (max.)
Figure B-6 Transverse Hall Generator HGT-3010, HGT-3030, and HGCT-3020 Dimensions
Table B-1 Cryogenic Hall Generator Specifications
Cryogenic HGCA-3020
HGCT-3020
Description
Cryogenic axial; phenolic package
Cryogenic transverse; ceramic package
Active area
0.030 inch diameter circle
0.040 inch diameter circle
Input resistance
1 ohm
1 ohm
Output resistance
1 ohm
1 ohm
Nominal control current (I
CN
)
100 mA
100 mA
Maximum continuous current
(non-heat sinked)
300 mA
300 mA
Magnetic sensitivity (I
C
= nominal control
current)
0.56 to 1.04 mV/kG
0.56 to 1.04 mV/kG
Maximum linearity error
(sensitivity vs field)
±1.0% RDG (-30 to +30 kG)
±2.0% RDG (-150 to +150 kG)
±1.0% RDG (-30 to +30 kG)
±2.0% RDG (-150 to +150 kG)
Zero field offset voltage (I
C
= nominal
control current)
±200 µV (max.)
±200 µV (max.)
Operating temperature range
4.2 K to 375 K
4.2 K to 375 K
Mean temperature coefficient of magnetic
sensitivity
±0.01%/K (approx.)
±0.01%/K (approx.)
Mean temperature coefficient of offset (I
C
= nominal control current)
±0.4 µV/K (max.)
±0.4 µV/K (max.)
Mean temperature coefficient of resistance ±0.6%/K (max.)
±0.6%/K (max.)
Leads
34 AWG copper w/Teflon insulation
34 AWG copper w/Teflon insulation
Содержание 420
Страница 6: ...Lake Shore Model 420 Gaussmeter User s Manual iv Table of Contents This Page Intentionally Left Blank ...
Страница 14: ...Lake Shore Model 420 Gaussmeter User s Manual 2 4 Installation This Page Intentionally Left Blank ...
Страница 22: ...Lake Shore Model 420 Gaussmeter User s Manual 3 8 Operation This Page Intentionally Left Blank ...
Страница 31: ...Lake Shore Model 420 Gaussmeter User s Manual Remote Operation 4 9 This Page Intentionally Left Blank ...
Страница 32: ......
Страница 46: ...Lake Shore Model 420 Gaussmeter User s Manual A 2 Appendix A This Page Intentionally Left Blank ...