3.0 GENERAL
This chapter covers aspects of Model 420 operation: Front Panel Controls (Paragraph 3.1), Front Panel
Functions (Paragraphs 3.2 thru 3.10), Default Settings (Paragraph 3.11), and Probe Handling Considerations
(Paragraph 3.12). See Chapter 4 for remote serial interface operation.
3.1 DEFINITION OF FRONT PANEL CONTROLS
The front panel consists of two major sections: 8 front panel keys (Paragraph 3.1.1), and a 2 row by 16
character Liquid Crystal Display (LCD, Paragraph 3.1.2).
3.1.1
Front Panel Keypad Definitions
AC / DC
Inter-
face
Range
Zero
Probe
Max
Hold
Max
Reset
420 Gaussmeter
LakeShore
Gauss /
Tesla
Relative
C-420-U-1-1
Figure 3-1 Model 420 Front Panel
Max Reset:
Works with Max Hold function. Returns Max reading to normal field reading. See Paragraph 3.2.
Max Hold:
Turns Max Hold feature ON and OFF. Captures and displays the highest field reading. Use Max
Reset key to clear reading. See Paragraph 3.2.
Zero Probe:
Zeros or nulls effects of ambient low level fields from the probe. To use, place tip of probe into
Zero Gauss Chamber and press
Zero Probe
. Display returns to normal readings in about 5 seconds. See
Paragraph 3.3.
Range:
Selects manual field measurement range or Autorange. Available ranges depend on which probe is
installed. There is a slight delay until the unit settles on the proper range. See Paragraph 3.4.
AC/DC:
Selects periodic (AC) or static (DC) magnetic fields. The AC selection provides RMS readings. See
Paragraph 3.5.
Gauss/Tesla:
Changes display units from gauss to tesla. Gauss (G) is used in the cgs system, where
1 G = 10
–4
T. Tesla (T) is used in the SI system, where 1 T = 10
4
G. See Paragraph 3.6.
Interface:
Adjusts the baud rate to 300, 1200, or 9600 during serial interface. See Paragraph 3.7.
Relative:
Displays the positive or negative deviation from setpoint in the lower line of the display. Often used
to offset large magnetic fields. May also be used with Max Hold and Alarm. See Paragraph 3.8.
Содержание 420
Страница 6: ...Lake Shore Model 420 Gaussmeter User s Manual iv Table of Contents This Page Intentionally Left Blank ...
Страница 14: ...Lake Shore Model 420 Gaussmeter User s Manual 2 4 Installation This Page Intentionally Left Blank ...
Страница 22: ...Lake Shore Model 420 Gaussmeter User s Manual 3 8 Operation This Page Intentionally Left Blank ...
Страница 31: ...Lake Shore Model 420 Gaussmeter User s Manual Remote Operation 4 9 This Page Intentionally Left Blank ...
Страница 32: ......
Страница 46: ...Lake Shore Model 420 Gaussmeter User s Manual A 2 Appendix A This Page Intentionally Left Blank ...