English
9
To remove the Canister, proceed as follows:
1
Disconnect the Filter Tubing from the blue
Vacuum Inlet (see section 4.2 regarding
handling instructions for the Aerosol Filter).
2
To release the Canister Holder, press down the
Release Arm while sliding the Holder towards
you. Remove the Canister from the Holder.
3
Release the Patient Suction Tubing (if
applicable).
4
Safely dispose of the contents of the Canister
according to local protocols.
5
Clean the LSU according to section 6.2 and
its reusable parts according to section 4.6.
4.3.4 Assembly of the Reusable Canister System
1
Place the yellow Gasket in the Lid, make sure
that the gasket is placed flat in the Lid and
not twisted.
2
If the Float Ball has been removed, hold the
lid upside down and snap the ball into place
as shown.
Note:
Never use the LSU without the Float Ball.
3
Secure the Lid to the Canister by turning it
clockwise until it is securely fastened.