laerdal LSU Скачать руководство пользователя страница 16

6.3 Disinfection of reusable parts

To disinfect the parts, the following methods are recommended:

Method

Temperature / Concentration

Contact Time

Post-treatment

Glutaraldehyde

Room temperature / Conc.: 2% 60 minutes

Rinse all parts warm water. 
Allow to dry.

Sodium Hypochlorite 
(not cleared for use in 
the US)

Room temperature / Conc.: 0.5% 20 minutes

Rinse all parts warm water. 
Allow to dry.

Virkon

Room temperature / Conc.: 1% 10 minutes

Rinse all parts warm water. 
Allow to dry.

Steam autoclaving

Autoclave at max. 121° C

60 minutes

Allow parts to cool.

English

16

6. Maintenance

6.1 General

•  Make sure the maintenance instructions  

under “After Use” are followed.

•  If the device is not in frequent use (i.e. less 

than once a month), the Device Test (see 
section 7) should be performed both on a 
monthly basis and after each use.

6.2 Clean the Cabinet and parts

Warning: Disconnect the LSU from external 
power prior to cleaning.

Use a minimum of liquid to prevent any electrical 
shock hazard. Do not immerse the LSU or allow 
it to stand in water or other liquids. This can 
damage the device, and cause electrical shock 
resulting in injury to persons.

•  Use a cloth or sponge that is dampened with  

a mild detergent (hand dishwashing liquid or  
similar) to clean the external surfaces of the  
LSU. Use a detergent that is compatible with  
the materials listed under section 11.7, and  
follow the detergent manufacturer´s 
instructions.

•  Use a cloth or sponge dampened with water  

and wipe the surfaces again.

•  Dry the surfaces using a clean cloth or a 

paper towel.

6.4 Charge the Battery

The internal rechargeable battery can be charged 
directly from external AC or DC power as 
described in section 3.4. No external charger is 
needed. To charge the battery, proceed as follows:

Make sure the Operating Knob is set to “0”.  
Connect either external AC or DC power to  
the LSU and charging will start automatically.

During charging the Battery Status Indicator  
will indicate approx. achieved battery capacity 
(see section 3.2.5 for details).

The LSU should be placed on charge for a 
minimum of 24 hours to reach full battery 
capacity. The fast charging gives approx. 80% 
capacity after 3 hours (for a new battery). 
Repetitive charging for only 3 hours is not 
recommended.

Содержание LSU

Страница 1: ...Laerdal Suction Unit Cat No 78 00 00 Reusable Cat No 78 00 30 Serres Directions for Use LSU...

Страница 2: ......

Страница 3: ...Clean the Cabinet and parts 16 6 3 Disinfection of reusable parts 16 6 4 Charge the Battery 16 6 5 External Battery Charger 17 6 6 Check the Battery Quality 17 6 7 Replace the Battery 17 6 8 Fastening...

Страница 4: ...pump any cleaning solution or other liquids through the vacuum pump i e through the Vacuum Connector This can damage the LSU Use only accessories supplied by Laerdal Medical or one of our authorised...

Страница 5: ...d If lit turn the LSU OFF and then ON again to check if the indication disappears Provided that the indication does not occur again the LSU can be operated If the Indicator continues to lit after thre...

Страница 6: ...n 7 3 4 Power Sources The LSU is delivered with an AC Power Cord for connection to AC mains and a DC Power Cord for connection to DC mains The device can be operated from the internal battery and can...

Страница 7: ...Suction Catheter Adapter English 7 The LSU Reusable Canister System consists of A 1000 ml transparent plastic Canister A Canister Lid w Float Ball An Aerosol Filter with tubing The Canister Lid conta...

Страница 8: ...wer as described in section 3 4 If you decide to operate the LSU from internal battery check that the battery is installed Check that the Patient SuctionTubing is securely connected to the PATIENT Inl...

Страница 9: ...cable 4 Safely dispose of the contents of the Canister according to local protocols 5 Clean the LSU according to section 6 2 and its reusable parts according to section 4 6 4 3 4 Assembly of the Reusa...

Страница 10: ...ing to the blueVacuum Inlet on the LSU Color code Blue on Blue 8 Insert the Canister in the canister holder and slide it into the LSU 9 Connect the yellow Angled Connector to the yellowVacuum Seal 10...

Страница 11: ...contamination of Laerdal Reusable Canister 4 6 1 Parts to be Decontaminated After Each Use 1 Canister Lid 2 Float Ball 3 Gasket 4 Vacuum Seal 5 Angled Connectors 6 Canister 7 Patient SuctionTubing 8 S...

Страница 12: ...osable Parts Suction bag with filter Serres SuctionTube Suction Bag with Filter Serres SuctionTube Reusable Parts Parts to be cleaned or replaced regulary Canister SerresVacuum Connector SerresVacuum...

Страница 13: ...parts are assembled correctly 5 3 2 Using the LSU Note The LSU should be operated and transported in the upright position to prevent overflow of suctioned material 1 Unwrap the Serres SuctionTube 2 C...

Страница 14: ...e Serres canister if needed according manufacturers instructions for use Note Clean the SerresVacuum Connector by wiping with a damp cloth or a sponge Do not autoclave or attempt to disassemble The Se...

Страница 15: ...the bag is fully inflated 10 Turn off the LSU 11 Fasten the SuctionTubing on the Suction Tubing Holder 12 Place the LSU on charge see section 6 4 5 4 Serres Suction Bag System Quick Reference Correct...

Страница 16: ...n water or other liquids This can damage the device and cause electrical shock resulting in injury to persons Use a cloth or sponge that is dampened with a mild detergent hand dishwashing liquid or si...

Страница 17: ...ged by the LSU and must be replaced We recommend charging a spare battery every 6 months when stored in room temperature at 25 C 77 F 6 5 External Battery Charger The LSU battery can also be charged i...

Страница 18: ...LSU needs service The program runs 4 different tests 1 It tests for occlusions in the Suction System including canister and tubing 2 It tests the vacuum build up efficacy of the Pump System How much v...

Страница 19: ...on takes approx one second fully block the Patient SuctionTubing 3 Keep the tubing blocked while LED 2 3 and 4 lights up 4 Release the tubing when LED 1 comes on again 5 Evaluate the test results See...

Страница 20: ...mmHg FAIL Check Connectors Tubes and Canister Lid for leakage or damage Check exhaust outlet for occlusion and run the Device Test again Test 3 The maximum achievable vacuum level of the LSU reaches t...

Страница 21: ...the battery see section 6 7 The LSU operates but little or no suction available Vacuum Seal blocked by float ball Unplug theVacuum Seal to release the vacuum Canister full Remove and replace the Canis...

Страница 22: ...et 78 04 30 LSU Reusable Canister Holder Serres 57 151 Serres Suction Bag 1000 ml Blue 58 23 181 Serres SuctionTube non sterile CH25 180 cm 78 12 06 SerresVacuum Connector 78 04 12 Disposable patient...

Страница 23: ...water bottle 79 35 00 Water Container Accessories all versions 78 20 00 Carrying Bag full covering 78 26 00 Wall Bracket w DC Power Cord 78 26 10 Wall Bracket w AC Power Cord US 78 26 20 Wall Bracket...

Страница 24: ...attery operation time free air flow at different settings 10 mmHg 80 120 200 350 500 min 3h20 2h20 1h30 1h 45 Approx noise levels free air flow at different settings mmHg 80 120 200 350 500 dBA 48 48...

Страница 25: ...n Catheter Adapter PC Suction Catheter Adapter Holder PC LSU Reusable Canister PC HT Lid w FloaterValve Cylinder Gasket for Lid LSU Reusable Canister Silicone Full covering Carrying Bag PVC coated Pol...

Страница 26: ...al Requirements for Safety CAN CSA C22 2 No 601 1S1 94 Supplement No 1 94 to CAN CSA C22 2 No 601 1 M90 Medical Electrical Equipment Part 1 General Requirements for Safety US Standards UL Standard No...

Страница 27: ...UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 se...

Страница 28: ...ng symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection f...

Страница 29: ......

Страница 30: ......

Страница 31: ......

Страница 32: ...8179 Rev I...

Отзывы: