laerdal LSU Скачать руководство пользователя страница 6

English

6

Typical remaining Battery capacity

Accuracy is dependent upon the lifetime and
condition of the battery, as well as, the condition
of the unit. Outside variables such as temperature
may also affect accuracy.

Battery Capacity

Output

> 50 %

4 LEDs lit.

30 - 50%

3 LEDs lit.

20 - 30%

2 LEDs lit.

< 20%

1 LED lit.

Battery Low; (min. 1 minute of operation left 
based on 500+ mmHg/free flow). 1 LED flashes.

During charging: indicates approx. achieved 
battery capacity.

Capacity

Output*

< 75%

The LED’s will be lit sequentially

75 - 80%

3rd LED’s lit and 4th flashing

> 80%

4 LED’s lit.

*Note:

 If no battery is installed, the battery status 

indicator will be fully lit for approx. 5 sec. until it 
is turned off.

During the Device Test

Indicates which step of the test is currently in 
progress or which corresponding test result is 
being displayed:
LED 1 (lower LED) lit = STEP 1, 
LED 2 lit = STEP 2 etc.

3.3 TEST-Button    

                     

This button allows you to run a 
4-step user initiated Device Test 
program to identify whether the 
LSU is operating satisfactorily or if it needs 
service (see section 7).

3.4 Power Sources

The LSU is delivered with an AC Power Cord for 
connection to AC mains and a DC Power Cord 
for connection to DC mains.
The device can be operated from the internal 
battery, and can be operated or charged from 
one of the following external Power Sources:
•  AC mains when used with the AC Power  

Cord: 100-240 VAC (50 / 60 Hz).

•  DC mains when used with the DC Power  

Cord:  12-28 VDC.

A Wall Bracket to hold the LSU during operation 
and (optional) charging is available separately 
(see section 10).

AC Connector Socket

DC Connector Socket

External power connection will cause  
the  External  Power  Indicator    to  be  lit.                      
Always check that it is lit when 
connecting external power or 
inserting the LSU into the Wall Bracket. 

3.5 Automatic Power-save Function

The LSU has an Automatic Power-
save Function that switches the pump 
motor off. 
             
While in this mode, the Power ON Indicator will 
flash slowly (approx. once per second).

This function will be activated when the 
Operating Knob is set to 200, 350 or 500+ 
mmHg and the actual vacuum level has been 
continuously higher than 120 mmHg for more 
than 2 minutes.
To exit the Power-save Mode and revert to 
normal operation, set the Operating Knob to 
any other position and then go back to required 
setting.

Содержание LSU

Страница 1: ...Laerdal Suction Unit Cat No 78 00 00 Reusable Cat No 78 00 30 Serres Directions for Use LSU...

Страница 2: ......

Страница 3: ...Clean the Cabinet and parts 16 6 3 Disinfection of reusable parts 16 6 4 Charge the Battery 16 6 5 External Battery Charger 17 6 6 Check the Battery Quality 17 6 7 Replace the Battery 17 6 8 Fastening...

Страница 4: ...pump any cleaning solution or other liquids through the vacuum pump i e through the Vacuum Connector This can damage the LSU Use only accessories supplied by Laerdal Medical or one of our authorised...

Страница 5: ...d If lit turn the LSU OFF and then ON again to check if the indication disappears Provided that the indication does not occur again the LSU can be operated If the Indicator continues to lit after thre...

Страница 6: ...n 7 3 4 Power Sources The LSU is delivered with an AC Power Cord for connection to AC mains and a DC Power Cord for connection to DC mains The device can be operated from the internal battery and can...

Страница 7: ...Suction Catheter Adapter English 7 The LSU Reusable Canister System consists of A 1000 ml transparent plastic Canister A Canister Lid w Float Ball An Aerosol Filter with tubing The Canister Lid conta...

Страница 8: ...wer as described in section 3 4 If you decide to operate the LSU from internal battery check that the battery is installed Check that the Patient SuctionTubing is securely connected to the PATIENT Inl...

Страница 9: ...cable 4 Safely dispose of the contents of the Canister according to local protocols 5 Clean the LSU according to section 6 2 and its reusable parts according to section 4 6 4 3 4 Assembly of the Reusa...

Страница 10: ...ing to the blueVacuum Inlet on the LSU Color code Blue on Blue 8 Insert the Canister in the canister holder and slide it into the LSU 9 Connect the yellow Angled Connector to the yellowVacuum Seal 10...

Страница 11: ...contamination of Laerdal Reusable Canister 4 6 1 Parts to be Decontaminated After Each Use 1 Canister Lid 2 Float Ball 3 Gasket 4 Vacuum Seal 5 Angled Connectors 6 Canister 7 Patient SuctionTubing 8 S...

Страница 12: ...osable Parts Suction bag with filter Serres SuctionTube Suction Bag with Filter Serres SuctionTube Reusable Parts Parts to be cleaned or replaced regulary Canister SerresVacuum Connector SerresVacuum...

Страница 13: ...parts are assembled correctly 5 3 2 Using the LSU Note The LSU should be operated and transported in the upright position to prevent overflow of suctioned material 1 Unwrap the Serres SuctionTube 2 C...

Страница 14: ...e Serres canister if needed according manufacturers instructions for use Note Clean the SerresVacuum Connector by wiping with a damp cloth or a sponge Do not autoclave or attempt to disassemble The Se...

Страница 15: ...the bag is fully inflated 10 Turn off the LSU 11 Fasten the SuctionTubing on the Suction Tubing Holder 12 Place the LSU on charge see section 6 4 5 4 Serres Suction Bag System Quick Reference Correct...

Страница 16: ...n water or other liquids This can damage the device and cause electrical shock resulting in injury to persons Use a cloth or sponge that is dampened with a mild detergent hand dishwashing liquid or si...

Страница 17: ...ged by the LSU and must be replaced We recommend charging a spare battery every 6 months when stored in room temperature at 25 C 77 F 6 5 External Battery Charger The LSU battery can also be charged i...

Страница 18: ...LSU needs service The program runs 4 different tests 1 It tests for occlusions in the Suction System including canister and tubing 2 It tests the vacuum build up efficacy of the Pump System How much v...

Страница 19: ...on takes approx one second fully block the Patient SuctionTubing 3 Keep the tubing blocked while LED 2 3 and 4 lights up 4 Release the tubing when LED 1 comes on again 5 Evaluate the test results See...

Страница 20: ...mmHg FAIL Check Connectors Tubes and Canister Lid for leakage or damage Check exhaust outlet for occlusion and run the Device Test again Test 3 The maximum achievable vacuum level of the LSU reaches t...

Страница 21: ...the battery see section 6 7 The LSU operates but little or no suction available Vacuum Seal blocked by float ball Unplug theVacuum Seal to release the vacuum Canister full Remove and replace the Canis...

Страница 22: ...et 78 04 30 LSU Reusable Canister Holder Serres 57 151 Serres Suction Bag 1000 ml Blue 58 23 181 Serres SuctionTube non sterile CH25 180 cm 78 12 06 SerresVacuum Connector 78 04 12 Disposable patient...

Страница 23: ...water bottle 79 35 00 Water Container Accessories all versions 78 20 00 Carrying Bag full covering 78 26 00 Wall Bracket w DC Power Cord 78 26 10 Wall Bracket w AC Power Cord US 78 26 20 Wall Bracket...

Страница 24: ...attery operation time free air flow at different settings 10 mmHg 80 120 200 350 500 min 3h20 2h20 1h30 1h 45 Approx noise levels free air flow at different settings mmHg 80 120 200 350 500 dBA 48 48...

Страница 25: ...n Catheter Adapter PC Suction Catheter Adapter Holder PC LSU Reusable Canister PC HT Lid w FloaterValve Cylinder Gasket for Lid LSU Reusable Canister Silicone Full covering Carrying Bag PVC coated Pol...

Страница 26: ...al Requirements for Safety CAN CSA C22 2 No 601 1S1 94 Supplement No 1 94 to CAN CSA C22 2 No 601 1 M90 Medical Electrical Equipment Part 1 General Requirements for Safety US Standards UL Standard No...

Страница 27: ...UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 se...

Страница 28: ...ng symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection f...

Страница 29: ......

Страница 30: ......

Страница 31: ......

Страница 32: ...8179 Rev I...

Отзывы: