6.8 Fastening brackets
The fastening bracket is used to hold the LSU in
the optional wall bracket. Inspect the fastening
brackets for wear and tear regularly.
Replace if worn. See section 10.
Fastening
brackets
6.9 Release arm for Canister holder
An extra release arm for the canister holder can
be attached. See section 10.
7 Device test
The Device test is a user initiated test program
for all models which can identify whether the
LSU operates satisfactorily or if the LSU needs
service.
The program runs 4 different tests:
1
It tests for occlusions in the Suction System
(including canister and tubing).
2
It tests the vacuum build-up efficacy of the
Pump System (How much vacuum is built up
within 3 seconds).
3
It tests the maximum achievable vacuum
level of the LSU (reaches this level within 10
seconds).
4
It tests if there are air leakages in the Pump
System (including canister and tubing).
7.1 Setup
Before the test can be performed make sure that:
• The Patient Suction Tubing is unwound and
not blocked or bent.
• The Suction Catheter Adapter is removed
from its holder (if applicable).
• The canister lid, T-bar, angled connector and
tubing are securely fastened.