7.2 Run the Test
Note:
If you need to interrupt the test and revert
to normal operation, turn the Operating Knob
to another position and then select the required
setting.
1
Press and hold the TEST-Button while setting
the Operating Knob to 500+ mmHg.
Note: Do not release the TEST-Button until min.
2 seconds after the Operating Knob has been set
to 500+ mmHg. The test will start immediately
2
As soon as LED 2 from the bottom of the
Battery Status Indicator comes on (takes
approx. one second), fully block the Patient
Suction Tubing.*
3
Keep the tubing blocked while LED 2 , 3 and
4 lights up.
4
Release the tubing when LED 1 comes on
again.
5
Evaluate the test results (See section 7.3)
6
After evaluating the test results turn the
Operating Knob to “0” to exit the Device Test.
Note:
To indicate that the LSU is in test mode,
the Power ON Indicator will flash rapidly (approx.
twice per second) until you exit the test program.
* If the tubing is not occluded within 2 minutes,
the test will be interrupted and the Power ON
Indicator will start to flash slowly (approx. once
per second). To restart the test, set the Operating
Knop to “0” and start over again.