5.3.1. Classic Test and Report
The
Calibration Report
shows:
✦
The Chromaticity Diagram (color gamut) of the monitor
compared to the sRGB, AdobeRGB and ISO-coated gam-
ut.
✦
The average and maximum Delta-Es of the measured color
patches
These Delta-E values are calculated with the CIE94 method and
therefore express perceptive color differences. The lower these
Delta-E values, the better. If these Delta-Es are below 2, the
values will be displayed in Green.
✦
The difference between the measured and targeted Gam-
ma, White Point Temperature and Luminance and their de-
viation in percentage.
The combination of these three elements allows you to have an
overall appreciation of the quality of the calibration: if the de-
viation between target values and current values is below 5%
it will be displayed in green. Differences exceeding 5% will be
highlighted in orange.