15
12.
Non-Volatile Memory Setting
If the non-volatile memory function is active (ON), the unit will run to the originally temperature,
speed and time program when the external power is recovered after a power failure.
12.1 To set the non-volatile memory function
, first press the, then press the Increase button to “8”, next
press the "Set/Confirm" button and the screen will display the character ON or OFF. Press the
"Increase" or "Decrease" button to change the parameter and press the "Set/Confirm" button to
save the change and exit.
12.2 Press the Timer
button, if the symbol “A” is displayed on the lower left corner of the screen, it
means the non-volatile memory function has been active.
13.
Setting of communication address for RS485 connection (Opt)
If the RS-485 communication kit is equipped on this product, please follow the instructions below
for address setting.
13.1 Press the Change/Confirm button, then press the Increase button to "9", press the
Change/Confirm button again to enter the "communication" address setting page.
13.2 The screen displays the current communication address, and use Increase or Decrease button to
change the address, within the range of 0 to 63
13.3 Press the Change/Confirm button to save and exit
9
Set Communication
Address
Exit
8
Non-volatile
memory function
On/Off
Exit