16 

 

14. 

Sum Run Time Checking 

The total run time of current operation can be checked via following steps. Run time starts 
counting once "Start/Stop" button is pressed, and stops when pressed again. 

     

 

14.1    Press "Set/Confirm" button and press the "Increase"    button to "10", then press "Set/Confirm" 

button again to see "Sum Run Time". 

14.2   This is the accumulated time of current operation, which cannot be changed. 

   

 

 

 

 

 

 

 

 

 

 

 

 

 
 
 

15. 

Operation and Switch off 

 
15.1 When all the above settings are done, press the Start/Stop button and the equipment will run 

according to the stored settings. 

15.2 When during operation, the Start/Stop button is pressed, the platform stops from shaking 

temporarily. The operation time remains on hold. 

15.3 Press the Start/Hold button again to resume operation, the pre-set time starts to count down from 

the remaining time when the operation was stopped. 

15.4 When during operation, if the Start/Stop button is held on, within 3 seconds the remaining 

operating time will be cleared to zero. Press the Start/Stop button once more, and the unit starts 
again to count down from the preset operation time. 

15.5 While the unit is in use, the current remaining operating time cannot be changed. If however 

changed at this time, it is invalid with the current operation. Only when the current operation has 
passed or stopped according to the above method, only if the instrument is re- started the new 
changed value will be effective.   

15.6 The equipment can be turned off by holding on the Power button on the control panel for 2 

seconds. At this time, the control board is still connected to the power supply, so the main switch 
on the right side of the equipment must be shut off to end operation completely. 

          

   

Special Attention! 
 

The  communication  address  is  used  as  an  identifying  code  for 
each  unit  to  pair  with  the  software  of  the  PC  terminal.  Each 
communication code must be connected to only one unit. Different 
communication addresses must be used if multiple units are to be 
connected to the same PC terminal.

 

 

10 

Check Sum Run Time 

Exit 

Содержание ZWY-111B

Страница 1: ...User s Manual This Manual applies to Modes ZWY 111B ZWY 211B ZWY 111G ZWY 211G ZWY 111C ZWY 211C ZWY 111D ZWY 211D ZWF 111 and ZWF 211 Economic Horizontal Shaking Incubator...

Страница 2: ...agriculture It is widely used for cell culturing hybridization cell aeration and solubility studies Etc Since the date of your purchase of this product after sale service will always be close to you t...

Страница 3: ...r is needed 6 A surge protector is recommended to avoid power related faults 7 In case of malfunction or burning smell the unit must be unplugged immediately Use a circuit breaker to cut off the power...

Страница 4: ...orm is still moving do not open the lid too much or too long as this might affect the temperature inside 8 The incubator must be kept away from volatile flammable explosive liquids or gases 9 Please k...

Страница 5: ...__ 10 9 Temperature Speed Alarm Settings _____________________________ 11 10 Refrigeration Defrost Settings ________________________________ 12 11 Temperature Calibration______________________________...

Страница 6: ...r in refrigerated units only has an independent overload protection CFC free refrigerant which causes no damage to ozone layer Applies on ZWY 2 ZWF 2 models Large doublefold tempered glass window and...

Страница 7: ...64 82 82 66 100 ml 64 82 82 66 250 ml 40 45 52 45 500 ml 28 33 38 32 750 ml 23 24 26 23 1000 ml 15 15 22 15 2000 ml 11 12 11 3000 ml 8 11 11 5000 ml 6 6 6 Test tube 18ml 208 208 208 208 Inner dimensio...

Страница 8: ...n Hold this button for 2 seconds to turn the unit on off the standby mode Temperature button Press this button to display the actual temperature value and the set temperature in turns Speed button Pre...

Страница 9: ...amber is deviated over the preset alarm limit the indicator flashes and the alarm beeps Heating indicator When the heater is connected with power supply this indicator will be light This indicator wil...

Страница 10: ...eep unscrewing screwing the adjusting bolt on the front left corner with the spanner until the unit is elevated leveled and sits firmly on the four blocks 4 2 The shaking mechanism of all reciprocal s...

Страница 11: ...the current status of temperature retaining Keep Temp Mode function press Increase or Decrease to change the setting 6 2 Press Set confirm to exit to the home page 1 Display current control mode and t...

Страница 12: ...imer page press Increase or Decrease to change the value if necessary If Timer is set as 0 the unit will keep running continuously 7 2 Press Set confirm to exit to the home page Special Attention 1 Th...

Страница 13: ...ess Increase or Decrease to change the time press Set confirm to save and switch to year setting page press Increase or Decrease to set the year and press Set confirm again to save and switch to the m...

Страница 14: ...Increase or Decrease to change the value 9 3 Press Set confirm to exit to the home page Special Attention 1 Deviation alarms can only be activated when unit is running and once actual temperature or...

Страница 15: ...ss the Set Confirm button to confirm and store the changed value 10 5 Explanation for setting auto defrosting cycle a Defrost Timer is the time interval between two cycles For example if this is set a...

Страница 16: ...it will start manual defrost for set defrosting period To abort press button and hold for 3 seconds Minimum time interval between two manual defrost cycles is 1 min Attention 1 The defrosting cycle wi...

Страница 17: ...ture should have changed due to the calibration Waiting for another one hour to let the temperature stabilize again at 37 C and check the thermometer value again and calculate the new difference f if...

Страница 18: ...tton if the symbol A is displayed on the lower left corner of the screen it means the non volatile memory function has been active 13 Setting of communication address for RS485 connection Opt If the R...

Страница 19: ...on within 3 seconds the remaining operating time will be cleared to zero Press the Start Stop button once more and the unit starts again to count down from the preset operation time 15 5 While the uni...

Страница 20: ...through the opening 16 5 Use the Allen key to loosen the A central fixing nut 16 6 Use the 8 10 spanner to hold the B adjusting bolt as still in the meantime loosen the C safety nut with 17 19 spanne...

Страница 21: ...cold air is excessive Close a part of the ventilation opening The ventilation fan is broken Notify the distributor to replace the fan The heater does not work Notify the distributor to repair the heat...

Страница 22: ...he equipment if it does not work Notify the distributor The equipment causes a strange loud noise The equipment is not placed horizontally Adjust the left back foot to make the equipment stable The fi...

Страница 23: ...20 18 Electronic Control System...

Страница 24: ...e the parts free of charge Your distributor will provide professional workmanship and will repair the unit without any charge When this one year warranty is expired we will continue to provide you wit...