Performance Parameters 

 

Both  the  interior  and  exterior  are  made  of  robust  materials  for  lifetime  operations.  The  inner 

chamber  is  made  of  high  quality  #304  stainless  steel  sheets,  with  4  round  coved  corners.  All 

exposed edges are de-burred to insure no sharp edges. The exterior is cold rolled steel finished 

with powder coated polyurethane finish, which is resistant to most chemicals and easily cleaned 

with mild household detergents. 

 

The ZHICHENG shaking incubator

’s microprocessor control system uses fast responding PT 100 

sensors  which  commands  and  executes  a  special  control  algorithm  that  energizes  a  solid-state 

switch to supply power to the heaters. 

 

The control electronics are protected trough a circuit breaker that may trip at 110% of loading rate, 

but will trip within 1 second at 150% of load rating. 

 

The compressor (in refrigerated units only) has an independent overload protection. 

 

CFC-free refrigerant which causes no damage to ozone layer.(Applies on ZWY-2/ZWF-2 models) 

 

Large  doublefold  tempered  glass  window  and  flourecent  light,  provide  complete  visibility  of 

chamber interior. 

 

Models  ZWY-111D  and  ZWY-211D  are  equipped  with  foot  pedal  for  lid  opening  even  when 

carrying full loads in both hands. 

 

LCD display presents all actual and    (pre)set parameters.   

 

Electronic timer, from 0 

– 9999 Mins, automatic stop, audio/ visual alarm. 

 

Password protection against unauthorized change of parameters. 

 

Audible and visual alarms for motor temperature and set point deviations 

 

Non-violate memory for set point retention after a power interruption.   

 

“Long-Life” brushless AC motor creates a smooth, quiet and uniformed shaking motion 

 

High capacity for up to 5L flasks (except ZWY-111B/211B, ZWF-111/211) 

 

Option: unit equipped with RS-485 interface communication can be connected to local network so 

that remote data logging and control can be achieved via laptop/ PCs. 

 

 

 

 

 

 

 

 

 

 

Содержание ZWY-111B

Страница 1: ...User s Manual This Manual applies to Modes ZWY 111B ZWY 211B ZWY 111G ZWY 211G ZWY 111C ZWY 211C ZWY 111D ZWY 211D ZWF 111 and ZWF 211 Economic Horizontal Shaking Incubator...

Страница 2: ...agriculture It is widely used for cell culturing hybridization cell aeration and solubility studies Etc Since the date of your purchase of this product after sale service will always be close to you t...

Страница 3: ...r is needed 6 A surge protector is recommended to avoid power related faults 7 In case of malfunction or burning smell the unit must be unplugged immediately Use a circuit breaker to cut off the power...

Страница 4: ...orm is still moving do not open the lid too much or too long as this might affect the temperature inside 8 The incubator must be kept away from volatile flammable explosive liquids or gases 9 Please k...

Страница 5: ...__ 10 9 Temperature Speed Alarm Settings _____________________________ 11 10 Refrigeration Defrost Settings ________________________________ 12 11 Temperature Calibration______________________________...

Страница 6: ...r in refrigerated units only has an independent overload protection CFC free refrigerant which causes no damage to ozone layer Applies on ZWY 2 ZWF 2 models Large doublefold tempered glass window and...

Страница 7: ...64 82 82 66 100 ml 64 82 82 66 250 ml 40 45 52 45 500 ml 28 33 38 32 750 ml 23 24 26 23 1000 ml 15 15 22 15 2000 ml 11 12 11 3000 ml 8 11 11 5000 ml 6 6 6 Test tube 18ml 208 208 208 208 Inner dimensio...

Страница 8: ...n Hold this button for 2 seconds to turn the unit on off the standby mode Temperature button Press this button to display the actual temperature value and the set temperature in turns Speed button Pre...

Страница 9: ...amber is deviated over the preset alarm limit the indicator flashes and the alarm beeps Heating indicator When the heater is connected with power supply this indicator will be light This indicator wil...

Страница 10: ...eep unscrewing screwing the adjusting bolt on the front left corner with the spanner until the unit is elevated leveled and sits firmly on the four blocks 4 2 The shaking mechanism of all reciprocal s...

Страница 11: ...the current status of temperature retaining Keep Temp Mode function press Increase or Decrease to change the setting 6 2 Press Set confirm to exit to the home page 1 Display current control mode and t...

Страница 12: ...imer page press Increase or Decrease to change the value if necessary If Timer is set as 0 the unit will keep running continuously 7 2 Press Set confirm to exit to the home page Special Attention 1 Th...

Страница 13: ...ess Increase or Decrease to change the time press Set confirm to save and switch to year setting page press Increase or Decrease to set the year and press Set confirm again to save and switch to the m...

Страница 14: ...Increase or Decrease to change the value 9 3 Press Set confirm to exit to the home page Special Attention 1 Deviation alarms can only be activated when unit is running and once actual temperature or...

Страница 15: ...ss the Set Confirm button to confirm and store the changed value 10 5 Explanation for setting auto defrosting cycle a Defrost Timer is the time interval between two cycles For example if this is set a...

Страница 16: ...it will start manual defrost for set defrosting period To abort press button and hold for 3 seconds Minimum time interval between two manual defrost cycles is 1 min Attention 1 The defrosting cycle wi...

Страница 17: ...ture should have changed due to the calibration Waiting for another one hour to let the temperature stabilize again at 37 C and check the thermometer value again and calculate the new difference f if...

Страница 18: ...tton if the symbol A is displayed on the lower left corner of the screen it means the non volatile memory function has been active 13 Setting of communication address for RS485 connection Opt If the R...

Страница 19: ...on within 3 seconds the remaining operating time will be cleared to zero Press the Start Stop button once more and the unit starts again to count down from the preset operation time 15 5 While the uni...

Страница 20: ...through the opening 16 5 Use the Allen key to loosen the A central fixing nut 16 6 Use the 8 10 spanner to hold the B adjusting bolt as still in the meantime loosen the C safety nut with 17 19 spanne...

Страница 21: ...cold air is excessive Close a part of the ventilation opening The ventilation fan is broken Notify the distributor to replace the fan The heater does not work Notify the distributor to repair the heat...

Страница 22: ...he equipment if it does not work Notify the distributor The equipment causes a strange loud noise The equipment is not placed horizontally Adjust the left back foot to make the equipment stable The fi...

Страница 23: ...20 18 Electronic Control System...

Страница 24: ...e the parts free of charge Your distributor will provide professional workmanship and will repair the unit without any charge When this one year warranty is expired we will continue to provide you wit...

Отзывы: