Chapter 7 - Technical Information & Specifcations
Compliance With International Standards
49
Accuracy
·
for glucose concentrations
< 4,2 mmol/L (75 mg/dL), the percentage of results
0,28 mmol/L, + 0,56 mmol/L and + 0,83
mmol/L (+ 5 mg/dL, + 10 mg/dL and + 15 mg/dL) of
the reference values
·
for glucose concentrations
> 4,2 mmol/L (75 mg/dL), the percentage of results
5 %, + 10 %, + 15 % and + 20 % of the
reference values
Power Supply
Powerless
Glucose Units
mg/dL, mmol/L
Dimensions
104x211x11 mm (4.1x8.3x.4 in)
Weight
40 gram (1.4 oz)
Memory
Unlimited results
Memory is limited to smart mobile device memory
Average
7-day, 14-day or 30-day averaging
Display
Dario™ application on a smart mobile device screen
Function
Hypo warning: Set by user.
Glucose Meter/lancing Device
Operational Temperature
10°C-45°C (50°F-113°F)
Storage Temperature Strip
2°C-32°C (36°F-90°F)
Altitude
Up to 3,400 meters (11,155feet)
Test Strip
Operational Temperature
10°C-45°C (50°F-113°F)
Storage Temperature Strip
2°C-32°C (36°F-90°F)
Operational Altitude
Up to 3,400 meters (11,155feet)
Do not store near solvents and ionizing radiation source.
CoMplIanCe WITH InTernaTIonal STanDarDS
This meter meets the electromagnetic immunity requirements as per EN ISO
15197 Annex A. The chosen basis for electrostatic discharge immunity testing
was basic standard IEC 61000-4-2. In addition, it meets the electro-magnetic
emissions requirements as per EN 61326. Its electromagnetic emission is thus low.
Interference from other electrically driven equipment is not to be anticipated.