6
Part number 21729, Rev 4
The alarms that may be set by the user are
HIGH TEMP ALARM
LOW TEMP ALARM
Setting security access code (optional)
When delivered, the freezer has the password security value in the
OFF setting. This means that access to modes and settings other
than RUN can be addressed.
If you want to restrict access to all menus other than the RUN
menu, which is a read-only menu, then you may activate the
password security (PASSWRD SECURITY) by switching it to the
ON setting. This setting resides in the CONFIGURE menu.
Once you activate password security, you will automatically
acquire the factory default password, which is a four-digit
numerical code.
Default access code: 1 2 3 4
You may change the code itself by addressing the
SETTINGS menu.
Please note the two menus on the right, and remember that the
choice of whether or not a code is needed is made in the
CONFIGURE menu; but the code itself is edited in the SETTINGS
menu.
Setting security access ON of OFF
1. Navigate to the CONFIGURE mode on the SYSTEM level menu,
and then press ENTER.
This will take you to the CONFIGURE menu.
Scroll down to PASSWRD SECURITY OFF (the current status).
2. Use the UP and DOWN keys to change from
PASSWRD SECURITY OFF to
PASSWRD SECURITY ON,
And press ENTER.
You have now activated password protection
CONFIGURE MENU:
Ring-back delay
Password security
(
Status message
)
LCD contrast
Run battery test
Battery back up
Sample
SETTINGS MENU:
Set time
Set date
Access code
High alarm
Temperature set
Low alarm
SYSTEM NORMAL
˃
CONFIGURE MODE
CALIBRATION MODE
CONFIGURE MODE
ADJUST
˃
PASSWRD SECURITY ON
USING UP & DOWN
Содержание GM-ULT-1786