Prima OPH
6135000-795
PRIMA OPH
7
REF
Issue 1.6
Printed on: July, 2020
WARNING LABELS AND NOTES
Caution
Observe all warnings labels and notes!
If any label is missing on your instrument or has become illegible,
please contact us or one of our authorized representatives.
We will supply the missing labels.
Instrument label plate
The instrument label plate indicates the following:
• Name of the unit
• Cat. No.
• Rated voltage and current consumption
• Rated frequency range
• Serial number
• Safety compliance
• Brand name
Brightness Control
After the illumination has been switched on, you can continuously adjust
the brightness of the fiber illumination by turning the appropriate knob.
Red & Green LED
LED light shown here at front of swivel arm, always glow “GREEN” during
working position of the instrument. This LED will glow “RED” when
suspension arm is parked at uppermost position, i.e. instrument is not in
working position. This is an indication to the user that he has to bring the
arm down to use it again to LED became green again. Please note that
at parking position, illumination LED will be off to save LED life but at the
same time fans will be in operating condition to keep the unit cool & safe.
Balance setting
Turn the screw to adjust the balance setting of the suspension arm.
Ceiling and wall mount
Don’t use the mount to support yourself
Maximum load
When the surgical microscope are mounted on the suspension systems
the overall weight of the microscope including accessories and coupling
must not exceed a maximum value of 7.5kg. Please consult the user
manual for the suspension system about the admissible maximum load.
Accompanying Documents must be consulted
Catalog Number
Compliance to Medical Device Directive 93/42/EEC
Protective Earth