background image

Prima OPH

Power switch not pressed 

Defective instruments fuse 

Defective power cable 

Line power failure 

Contact in-house Technician

  

TROUBLESHOOTING TABLE

Insert light guide as far as it will
go 

Failure of electronics

6135000-795 

 PRIMA OPH 

29

       Issue 1.6

  Printed on: July, 2020

Chance light guide.

Insert light guide properly to get
Maximum illumination

Adjust brightness using the
brightness control knob

Insert light guide properly to get
maximum illumination

Contact service dept.

Change power cable

Change instrument fuse

Press power switch

Plug in power cable.

REMEDY 

PROBLEM

POSSIBLE CAUSE

No Illumination at all

Insufficient
Illumination

Surgical field
illumination
inoperative or blinks

Power cable not plugged

Failure of suspension
system electronics

Light guide not properly
inserted in arm or
microscope.

Light guide not properly
inserted in arm or
microscope.

Brightness level set too low

Defective light guide
(illumination not uniform)

Light guide not properly
inserted in microscope arm

Switch off via micro switch on
suspension system

Move suspension arm of
suspension system into working
position

Illumination surgical field using an 
or illuminator. Contact service dept.

This instruments a high grade technological product and not required any special periodical
Maintenance if handed carefully. To ensure optimum performance and safe working order of the
Instrument, its safety must be checked once every 12 months. We recommended having this check
performance by our service representatives as part of regular maintenance work. If a failure occurs
which you cannot correct using the trouble-shooting table, attach a sign to the instrument stating
out of order and contact our service representative for servicing part or circuit diagram etc..

Содержание Prima OPH

Страница 1: ...rument as well as to avoid injury while operating instrument Understanding this manual completely before use is highly recommended User Manual PRIMA OPH Operating Surgical Microscopy Part No 6135000 7...

Страница 2: ...RIC CONNECTION CONTROLS PRIMA WIRING CODING DIAGRAM INSTRUCTION FOR USING THE MICROSCOPE CHANGING THE OBJECTIVE EYEPIECES REPLACING THE ILLUMINATION SOURCE FUSE REPLACEMENT THERMAL CUT OFF TENSION ADJ...

Страница 3: ...es of this Microscope are 1 The observation head can easily be positioned with the help of suspension arm 2 An advanced 5 step magnification changer allows an optical magnification for a particular su...

Страница 4: ...al attention to these safety notes Safety is only ensures when this instrument is operated properly Please read this manual carefully before turning the instrument on Also read through the user s manu...

Страница 5: ...CONDUCTED TO ENSURE CONTINUED SAFE USE OF THIS INSTRUMENT WARNING ENSURE THAT THE VOLTAGE APPLIED TO THE UNIT IS THE SAME AS THE VOLTAGE THAT IS INDICATED ON THE DATA PLATE OTTHERWISE DAMAGE TO THE UN...

Страница 6: ...NSTRU MENT TO AVOID DAMAGE TO THE UNIT MAY OCCUR SEE THE CARE AND MAINTENANCE SECTION FOR DETAILED CLEANING INSTRUCTIONS CAUTION MEDICAL ELECTRICAL EQUIPMENT NEEDS SPECIAL PRECAUTIONS WITH RESPECT TO...

Страница 7: ...Constant Voltage stabilizer with this Instrument Do not force cable connections If the male and female parts do not readily connect make sure that they are appropriate for one another If any of the co...

Страница 8: ...The power cord being used is the one designed for use with this instrument BEFORE EVERY USE AFTER RE EQUIPPING THE INSTRUMENT Make sure that all Requirement for operation are fulfilled Go through the...

Страница 9: ...of the instrument This LED will glow RED when suspension arm is parked at uppermost position i e instrument is not in working position This is an indication to the user that he has to bring the arm d...

Страница 10: ...ct upright position of the transport package Keep Dry Transport package shall be kept away from rain The CE mark an acronym for the French Conformity Europeenne Certifies that a product has met Europe...

Страница 11: ...Prima OPH _ 6135000 795 PRIMA OPH 9 Issue 1 6 Printed on July 2020 WARNING LABELS AND NOTES continued...

Страница 12: ...mn 5 Suspension Arm 7 Suspension arm movement locking knob 9 Suspension arm hydraulic movement lock 11 Swivel arm cover 13 Magnichanger 15 Common Main Objective 17 Z axis 19 Foot paddles for Z axis 2...

Страница 13: ...PIECE 12 5X16 6135012 CMO 200 6133200 CMO 175 6133176 6134130 111 BAYONET MOUNT NIKON 6134130 110 BAYONET MOUNT CANAN 6134130 112 BAYONET MOUNT SONY NEX 5N 6134130 113 BAYONET MOUNT CANON EOSM M10 61...

Страница 14: ...6165000 973 36 612086 972 DESCRIPTION Swivel arm assembly Suspension arm assembly XY Coupling Z Axis Coupling 12 5x Eyepiece 10x Eyepiece 45 inclined Binocular Head 210 tiltable binocular tube 200mm...

Страница 15: ...se with brakes on castor wheels or the type of mounting system 2 Column Floor or Ceiling or Wall Mount as ordered 3 Swivel arm and suspension arm assembly with fibre optic cable 4 XYZ Assembly or Z Ax...

Страница 16: ...ly as shown in fig 3 4 Place the wheel support genteelly on the floor as shown in fig 4 5 Uplift the top cover as shown in fig 5 place it close to the wheel support assembly Kindly repeat the same pro...

Страница 17: ...Remove the 2 no s of Foam sheets as shown in fig 8 9 Uplift the Centre Support Assembly as shown in fig 9 10 Place the centre Support Assembly on the wheel support assembly by aligning Corresponding...

Страница 18: ...screws from the sides shown as C in fig 14 3 Make sure that during assembly of the column ring D is fixed in place as shown in fig 15 Retrieve the swivel arm and suspension arm assembly from the pack...

Страница 19: ...d within the arm Mounting the light guide Switch off the illumination system The light guide comes pre routed through swivel arm that houses the LED illumination system and the suspension arm Insert t...

Страница 20: ...ows and cooling fan starts running Keep the intensity control knob at Minimum level before switching on the system To save burning life of LED either switch OFF the microscope or raise the suspension...

Страница 21: ...Prima OPH 6135000 795 PRIMA OPH 19 PRIMA WIRING CODING DIAGRAM Issue 1 6 Printed on July 2020...

Страница 22: ...g screw 3 Lock the Up Down movement of suspension arm using locking knob B in fig 9 after coarse focusing of the attendance area 4 Adjust the eye distance as per IPD scale according to your convenienc...

Страница 23: ...center position XY movement is done through joystick provided foot control Z axis movement Z axis coupling allows fine Motorized focusing of the area of interest within a travel range of 40mm Two foo...

Страница 24: ...RIMA OPH Issue 1 6 Printed on July 2020 The ergonomic 45 or 210 degree binocu lar tube is provided for the main surgeon The large tilt range allows work with minimum fatigue The standard equipment inc...

Страница 25: ...in the eye tubes of observation head 3 A range of objectives eyepieces can be selected by choice Open the swivel arm cover s Detach the fibre optic cable and replace the illumination assembly A with...

Страница 26: ...ll start working again To replace LED user may call LABOMED service personal or authorized dealer LED Specification 3 7V 13 5 Amp After Supplementary accessories are mounted the additional load of sus...

Страница 27: ...he video cable from the video modules e g video monitor USB monitor and the camera control unit Release locks by pressing the upper parts of the locks downwards Bring the unit into its moving position...

Страница 28: ...r and contact our service representative for servic ing Part or circuit diagram etc CARE INSTRUCTIONS Keep accessories away from dust when not in use e g protect them from dust cover Remove dust with...

Страница 29: ...TION Observe work safety and health protection of persons responsible for processing contaminated products Current regulation of hospital hygiene and prevention of infection must be observed in the Pr...

Страница 30: ...ing medical devices The unit is marked with CE and is compliance to IEC 60601 1 2007 3rd edition Following disinfectants are recommended for cleaning and disinfection 1 Normal household bleach Sodium...

Страница 31: ...perly inserted in arm or microscope Light guide not properly inserted in arm or microscope Brightness level set too low Defective light guide illumination not uniform Light guide not properly inserted...

Страница 32: ...ed properly Brakes on wheels not used Friction adjustment screw on suspension system tightened too firmly Defective fan Failure of system electronics Thermal cut off in lamp housing is contaminated In...

Страница 33: ...DISPOSAL Prima OPH Disposal must comply with locally applicable laws regulations 6135000 795 PRIMA OPH 31 Issue 1 6 Printed on July 2020...

Страница 34: ...50 x 50 mm 40mm 6 Speed Foot Control Assistant Binocular attachment Double Beam Splitter 50 50 and 80 20 70 30 0 210 inclinable ergo tube CMO 175 fixed CMOS Adapter Floor stand 600mm width 620mm leng...

Страница 35: ...cified below The customer or user of the Prima OPH should ensure that it is used in such an environment RF Emissions CISPR 11 Harmonics IEC 61000 3 2 Flicker IED 61000 3 3 Complies Class A Group 1 Cla...

Страница 36: ...evel Electromagnetic Environment Guidance 95 Dip for 0 5 Cycle 60 Dip for 5 Cycles 30 Dip for 25 Cycles 95 Dip for 5 Seconds 95 Dip for 0 5 Cycle 60 Dip for 5 Cycles 30 Dip for 25 Cycles 95 Dip for 5...

Страница 37: ...mended separation distance in meters m Field strengths from fixed Rf transmitters as determined by an electromagnetic site survey should be less than the compliance levels in each frequency range Int...

Страница 38: ...paration Distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to...

Страница 39: ...525 600 525 600 150 150 150 150 135 135 Prima OPH DIMENSIONS 90 90 536 150 150 135 1280 90 250 525 600 1733 37 6135000 795 PRIMA OPH Ceiling Mount Wall Mount Floor Mount Issue 1 6 Printed on July 2020...

Страница 40: ...blue light and emits green Specifies the luminous flux from a light source onto a certain area The unit of measure for illumination intensity is Lux Lx Light Emitting Diode Electronic semiconductor d...

Страница 41: ...dental or consequent damages for any negligence breach of warranty strict liability or any other damages resulting from or relating to design manufacture sale use or handling of the product PRODUCT CH...

Страница 42: ...Change App By DCR March 22 2016 July 05 2017 June 15 2019 DCR 4A 17 DCR 20A 19 DCR 02 2020 S Bal Wiring Coding Diagram added UPS requirement added System Diagram updated as per marketing brochure Cle...

Страница 43: ...hone 510 445 1257 Fax 510 991 9862 Email sales laboamerica com www laboamerica com Labomed Europe Essebaan 50 NL 2908 LK Capelle a d IJssel The Netherlands Tel 31 0 10 4584222 Fax 31 0 10 4508251 E ma...

Отзывы: