CLEANING AND SERVICING
6135000-795
PRIMA OPH
Surgical Microscope Maintenance/Servicing Check
Issue 1.6
Printed on: July, 2020
CARE & MAINTENANCE (Fig.-11)
This instrument is a high grade technological product and not required any special periodical
Maintenance if handed carefully. To ensure optimum performance and safe working order of the
Instrument, its safe functioning must be checked once every 12 months as per table below. We
Recommended having this check performance by our service representative as part of regular
Maintenance work. If a failure occurs which you cannot correct using the trouble-shooting table,
Attach a sign to the instrument stating out of order and contact our service representative for servic-
ing Part or circuit diagram etc.
CARE INSTRUCTIONS:
• Keep accessories away from dust when not in use, e.g. protect them from dust cover.
• Remove dust with a pneumatic rubber bulb and soft rush.
• Use special optics cleaning cloths and pure alcohol for cleaning lenses and eyepieces.
• Protect your colposcope from moisture, fumes, acids and cosmetic materials. Do not store chemicals
close to the instrument.
• Protect it from improper handling. Never install other devices sockets or unscrew optical system and
Mechanical parts unless explicitly instructed to do so in this manual.
• Protect the microscope from oil and grease. Never oil or grease the guide surfaces or mechanical
parts.
• Remove coarse contamination using a damp disposable cloth.
• Use disinfectants based on the following ingredients: aldehyde, alcohols, quaternary ammonium
compounds.