Prima OPH
REMEDY
30
6135000-795
PRIMA OPH
Issue 1.6
Printed on: July, 2020
TROUBLESHOOTING TABLE (continued)
PROBLEM
POSSIBLE CAUSE
Up & Down motion
of screw on
suspension system
too stiff.
Microscope Unstable.
No image visible in
field of view.
Magnichanger is not
Indexed properly.
Brakes on wheels not
used.
Friction adjustment screw
on suspension system
tightened too firmly.
Defective fan. Failure of
system electronics.
Thermal cut-off in lamp
housing is contaminated.
Index magnichanger properly.
Use Brakes.
Loosen friction adjustment
screw on suspension system
as require.
Contact service dept.
Illumination surgical field using
an or illuminator. Contact
Service dept.
Clean thermal cut-off with a
dry brush; blow it clean, if
necessary.