Magna
Magna is intended for use in the electromagnetic environment specified below. The customer or
user of the Magna should ensure that it is used in such an environment.
MAGNA
GUIDANCE TABLES
Guidance and Manufacturer’s Declaration
Electromagnetic Immunity
Equipment and Systems that are NOT Life-Supporting
Guidance and manufacturer’s Declaration – Electromagnetic Immunity
Immunity Test IEC 60601 Test Level Compliance
Level
Electromagnetic environment
Guidance
Recommended Separation Distance:
d=(3.5/v1)(Sqrt P)
d=(3.5/E1)(Sqrt P)
80 to 800 MHz
d=(7/E1)(Sqrt P)
800 MHz to 2.7GHz
6129000-795
Issue 1.7
Printed on May,2020
58
Conducted RF
IEC 61000-4-6
AC Mains
Radiated RF
IEC 61000-4-3
Enclosure
3 V with 6V ISM,
home:6V amateur
80% AM at 1kHz or
risk frequency
150kHz - 80MHz
3 V with 6V ISM,
home:6V amateur
80% AM at 1kHz
or risk frequency
150kHz - 80MHz
3V/m, home: 10V/
m 80% AM at 1kHz
or risk frequency
80 MHz - 2700MHz
3V/m, home: 10V/
m 80% AM at
1kHz or risk
Frequency
80 MHz -
2700MHz
Note 1: At 80 MHz to 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
*Field strengths from fixed transmitters, such as base stations for radio(cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To access the electromagnetic
Environment due to fixed RF transmitters, an electromagnetic site survey should be considered. The measured field strength in the
location in which the ME Equipment or ME system should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the ME Equipment or ME System.
*Over the frequency range 150kHz to 80 MHz, field strengths should be less than [V1] V/m.
Portable and mobile RF communica-
tions equipment should be used no
closer to any part of the Magna, includ-
ing cables than the recommended sepa-
ration distance calculated from the
equation applicable to the frequency of
the transmitter.
Where P is the max output power rating of the
transmitter in watts (W) according to the
transmitter manufacture and d is the Recom-
mended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
Should be less than the compliance levels in
each frequency range.
Interference may occur in the vicinity of equip-
ment marked with the following symbol.