8A.1. LEVELING THE WALL MOUNT PLATE ASSEMBLY
It is important to ensure that wall mount (6137100-806) is level both horizontally and vertically,
after installation. Leveling of the wall mount assembly is necessary to prevent the microscope
system from drifting from side to side when it is being used at full extension.
1. Slightly tighten the bolts on the top side- leaving the bottom bolts slightly loosened.
2. Place a level across the top of the back plate to check the horizontal leveling. See Figure 11 .
3. Use the 17mm socket and ratchet to slightly loosen the 3/8” x 4" hex head lag screws in the
corners of the back plate.
4. Adjust the back plate until it is level and tighten the three lag screws on the bottom leaving the
top lag screws loosened slightly.
5. Place a level vertically along the face of the wall Mount, on both sides, to ensure it is perpendic-
ular to the floor. See Figure 11 .
6. If the wall mount is not perpendicular, insert shim material between the wall mount and the wall
next to the lag screws Refer figure-11.
Check with the level and tighten the lag screws once the mount is level.
7. Check the other side and insert shims as needed and then tighten the lag screws.
8. When the Wall Mount is level, both horizontally and vertically, securely tighten all Six lag
screws.
Shim
Level
Wall Mount
Assembly
Figure-28