5B. LEVELING THE CEILING MOUNT SYSTEM
It is important to ensure the column for the ceiling mount is Plumb after installation. The column must be
perfectly vertical to prevent the microscope system from drifting from side to side.Use following procedure:
1. Use the 10mm Allen key to slightly loosen all of the 6M10 bolts.
2. Place a level across the bottom of ceiling mount as shown in Figure and to check the Vertical leveling.
Use Shims as required and adjust the RAWL Bolts/ Lag bolts until the column is Vertical.
3. Tighten all screws securely.
NOTICE:Shims and Shim materials are to be provided by the customer and are not furnished by
Labomed. There are numerous shims and shim materials commercially available and no specific type or
brand name is required.
Heavy Duty vibration free Ceiling Mount
Normal Ceiling Mount
Figure-13
Figure-14
Joist
Master Plate
Shim
Column
Level
Wooden membrane
Wooden membrane
Joist
Master Plate
Shim
Column
Level