eVO 350
TROUBLESHOOTING
18
Chart of Common Errors
ISSUE
PROBABLE CAUSE
POSSIBLE SOLUTION
Lamp won’t turn on.
Incorrect input power supplied to
the eVO 350 Slit Lamp.
Check the outlet to ensure proper power in
being supplied.
Defective Power Cord.
Bulb may be blown out.
Replace the Power Cord.
Bulb not installed properly.
Bulb not installed properly.
Replace Bulb.
Defective Power Supply.
Replace the Power Supply.
Slit Lamp won’t move.
Rubber stopper may be attached
under the joystick.
Remove the rubber stopper.
Base Lock Screw may be tightened.
Loosen the Base Lock Screw.
Rough base
movement.
Rubber stopper may be attached
under the joystick.
Remove the rubber stopper.
Bearings may be damaged.
Replace the base.
Shaft may be damaged.
Replace the base.
Fixation light does not
light up.
Fixation Light Harness not plugged
into the Power Supply Assembly.
Ensure the Fixation Light Harness is properly
seated in the Power Supply Assembly.
Defective Power Supply.
Replace the Power Supply.
Light too dim.
Incorrect wattage for bulb being used. Replace with the proper Bulb.
with bulb housing.
Double slit visible in
microscope.
Microscope not focused on focusing
rod before use.
Install focusing rod and check to ensure
microscope is focused on it.
Check bulb and ensure notch lines up
Check bulb and ensure notch lines up
with bulb housing.
8126300-795
eVO 350
Issue 1.2
Printed on June,2020
The following is a checklist of items that need to be assessed in order to determine if the eVO 350 Slit Lamp
requires servicing.
•
Check the outside of the Slit Lamp for any damage or missing components.
•
Inspect the power cord for damage.
•
Test the lamp by turning the lamp on and turning the light all the way to it’s brightest setting, and
all the way down to its lowest setting.
•
Check to ensure all switches are functioning properly.
•
Check the filters by cycling through all the options.
•
Check the Slit Wheel by cycling through all the options.
•
Check the base movement.
The following chart outlines some common issues with the eVO 350 Slit Lamp and some steps you can
take to correct the issue. If problems persist, please contact the LABOMED as listed in the
Introduction
section of this manual.