CHANGING THE HALOGEN BULB
WARNING:
NEVER REMOVE A BULB THAT HAS
RECENTLY BEEN IN USE AS IT WILL BE VERY HOT.
WAIT UNTIL IT HAS COOLED AND USE GLOVES OR A
THICK CLOTH WHEN HANDLING ANY HALOGEN BULB.
WARNING
: NEVER TOUCH A HALOGEN BULB WITH
BARE HANDS AS FINGERPRINTS WILL SHORTEN THE
BULB LIFE.
1. Remove input power to the instrument.
2. Open the bulb door and pull out the bulb holder by
pulling the metal pins on the holder. Refer Figure 21.
3. Swing the retaining clamp from the bulb by loosen
the screw. Refer Figure 22.
4. Pull the bulb from the unit. Refer Figure 23. and
replace the bulb with the correct bulb as indicated in
the specifications section of this manual.
Note:
Position the Bulb so the Cut Out in the metal
collar of the Bulb lines up with the Notch
in the lamp housing. Refer to Figure-24.
7. Move the Retaining Clamp back into its original
position. Refer to Figure-22.
8. Place the bulb holder back on the position.
9. Close the bulb door.
Figure-21 Bulb Holder
Figure-22 Retaining Clamp
Figure-23 Bulb
Retaining
spring
Retaining
Spring
Rotated
Bulb
Holder
Bulb Holder
Metal Pins
Retaining Clamp
Screw
Bulb