|        LABNET PRISM MICROCENTRIFUGE

7.6 Selection of Operating Time, Momentary Operation, Start/Stop

Operating time can be selected from 0.5 min to 99 min by pressing the TIME button and adjusting with 
the control knob. The time can be set in 0.5 minute increments from 0 to 10 minutes and in 1 minute 
increments from 10 to 99 minutes. After 99 minutes, the display shows “--” which indicates continuous 
run. In this mode, the centrifuge will run until manually stopped. To start a run, press the control knob. 
When the preselected time expires, the centrifuge will stop automatically. To stop the centrifuge prior 
to the expiration of set time, press the control knob. The centrifuge may be operated for a short run by 
pressing and holding the control knob. The centrifuge will continue to run as long as the control knob is 
depressed and the time, in seconds will count up on the time display.

 8.0  Service and Maintenance

8.1 Centrifuge Service

The brushless motor in the Prism™ Microcentrifuge requires no routine maintenance. Any required 
service should be performed by authorized, qualified personnel only. Repairs performed by unauthorized 
personnel may void the warranty.

8.2 Cleaning

Always keep the centrifuge housing, rotor chamber, rotor and rotor accessories clean. All parts should be 
wiped down periodically with a soft cloth. For more thorough cleaning, use a neutral cleaning agent (pH 
between 6 and 8) applied with a soft cloth. Excessive amounts of liquid should be avoided. Liquid should 
not come into contact with the motor. After cleaning, ensure that all parts are dried thoroughly by hand 
or in a warm air cabinet (maximum temperature 50°C).

The rotor should be cleaned after each use. When spinning samples containing phenol or phenol 
chloroform, the rotor should be cleaned immediately after use.

Should a spill of infectious materials occur within the rotor or chamber, the unit should be disinfected. 
This should be performed by qualified personnel with proper protective equipment.

8.3 Replacing Fuses

Check the fuse when it is recommended in the Troubleshooting Guide located in this manual. The fuse 
holder is located in the power inlet on the rear of the unit. Disconnect the power cord from the power 
inlet. Open the fuse holder drawer by inserting a small screwdriver under the tab and prying it open. 
Remove the innermost (operative) fuse from its retaining tabs and replace the fuse if necessary. A spare 
fuse is located in the outermost chamber of the fuse drawer. Replace only with a fuse of exactly the 
same value as the original. Fuse type may be found in the Technical data section of this manual.

Содержание Prism

Страница 1: ...Prism Microcentrifuge Instruction Manual Catalog Numbers C2500 C2500 230V This manual is available in additional languages at www labnetlink com...

Страница 2: ...sories 3 6 2 Rotor Maintenance 4 6 3 Removing and Installing the Angle Rotor 4 6 4 Loading the Rotor 4 6 5 Overloading Rotors 4 7 0 Operation 5 7 1 Attaching Rotor Lid 5 7 2 Closing the Lid 5 7 3 Lid...

Страница 3: ...equipment Never centrifuge flammable explosive or corrosive materials Never centrifuge hazardous materials outside of a hood or proper containment facility Always load the rotor symmetrically Each tu...

Страница 4: ...nual 1 each Warranty card 1 each Power cord 1 each 5 0 Installation The centrifuge should be installed on a rigid even surface such as a stable laboratory bench countertop etc To guarantee sufficient...

Страница 5: ...ncluded Cat No AR110 Strip Spin Adapter Cat No C2400 SS Tubes Accepted 0 2 mL tubes or 8 x 0 2 mL strips Maximum Speed 15 000 rpm Centrifuge Radius Range 4 32 cm to 5 33 cm RCF G value Range 10 874 x...

Страница 6: ...otor wrench 6 4 Loading the Rotor Tubes to be loaded should be filled equally by eye The difference in the weight between the tubes should not exceed 0 1 gram Tubes should always be loaded so that the...

Страница 7: ...ARNING Do not attempt to open the lid of any centrifuge until the rotor has come to a complete stop In the event of a power failure or malfunction it may be necessary to open the lid manually 1 Discon...

Страница 8: ...ousing rotor chamber rotor and rotor accessories clean All parts should be wiped down periodically with a soft cloth For more thorough cleaning use a neutral cleaning agent pH between 6 and 8 applied...

Страница 9: ...and or time bAL Error Message Indicates imbalance Tubes not inserted symmetrically in rotor holes Load tubes symmetrically see Section 3 4 on Loading the rotor Sample liquid in tubes not balanced Make...

Страница 10: ...e has an automatic g force conversion program so g values are automatically calculated and can be displayed on the control panel for this centrifuging radius If adapters or smaller tubes are used the...

Страница 11: ...or contact your local support office Corning s Customer Service team will help arrange local service where available or coordinate a return authorization number and shipping instructions Products rec...

Страница 12: ...r local Corning sales office www labnetlink com For a listing of trademarks visit www corning com clstrademarks All other trademarks are the property of their respective owners 2021 Corning Incorporat...

Отзывы: