Prism™ Microcentrifuge
Instruction Manual
Catalog Numbers:
C2500, C2500-230V
This manual is available in additional languages at
www.labnetlink.com
.
Страница 1: ...Prism Microcentrifuge Instruction Manual Catalog Numbers C2500 C2500 230V This manual is available in additional languages at www labnetlink com...
Страница 2: ...sories 3 6 2 Rotor Maintenance 4 6 3 Removing and Installing the Angle Rotor 4 6 4 Loading the Rotor 4 6 5 Overloading Rotors 4 7 0 Operation 5 7 1 Attaching Rotor Lid 5 7 2 Closing the Lid 5 7 3 Lid...
Страница 3: ...equipment Never centrifuge flammable explosive or corrosive materials Never centrifuge hazardous materials outside of a hood or proper containment facility Always load the rotor symmetrically Each tu...
Страница 4: ...nual 1 each Warranty card 1 each Power cord 1 each 5 0 Installation The centrifuge should be installed on a rigid even surface such as a stable laboratory bench countertop etc To guarantee sufficient...
Страница 5: ...ncluded Cat No AR110 Strip Spin Adapter Cat No C2400 SS Tubes Accepted 0 2 mL tubes or 8 x 0 2 mL strips Maximum Speed 15 000 rpm Centrifuge Radius Range 4 32 cm to 5 33 cm RCF G value Range 10 874 x...
Страница 6: ...otor wrench 6 4 Loading the Rotor Tubes to be loaded should be filled equally by eye The difference in the weight between the tubes should not exceed 0 1 gram Tubes should always be loaded so that the...
Страница 7: ...ARNING Do not attempt to open the lid of any centrifuge until the rotor has come to a complete stop In the event of a power failure or malfunction it may be necessary to open the lid manually 1 Discon...
Страница 8: ...ousing rotor chamber rotor and rotor accessories clean All parts should be wiped down periodically with a soft cloth For more thorough cleaning use a neutral cleaning agent pH between 6 and 8 applied...
Страница 9: ...and or time bAL Error Message Indicates imbalance Tubes not inserted symmetrically in rotor holes Load tubes symmetrically see Section 3 4 on Loading the rotor Sample liquid in tubes not balanced Make...
Страница 10: ...e has an automatic g force conversion program so g values are automatically calculated and can be displayed on the control panel for this centrifuging radius If adapters or smaller tubes are used the...
Страница 11: ...or contact your local support office Corning s Customer Service team will help arrange local service where available or coordinate a return authorization number and shipping instructions Products rec...
Страница 12: ...r local Corning sales office www labnetlink com For a listing of trademarks visit www corning com clstrademarks All other trademarks are the property of their respective owners 2021 Corning Incorporat...