Labnet Prism Скачать руководство пользователя страница 6

4

 

         

|        LABNET PRISM MICROCENTRIFUGE

6.2 Rotor Maintenance

The rotor should be cleaned thoroughly after each use. 

Thorough cleaning must be performed when 

spinning samples containing phenol or phenol chloroform.

 Periodically inspect the rotor for dents, 

dings, scratches, discoloration and cracks. If any damage to the rotor is found, discontinue use of the 
rotor immediately and replace.

6.3 Removing and Installing the Angle Rotor

The Prism™ Microcentrifuge comes complete with a standard 24-place rotor installed. To remove 
the rotor for cleaning, remove the rotor securing screw from the motor shaft by turning the screw 
counterclockwise, using the rotor wrench. Lift the rotor directly upward in a straight vertical motion.

Figure 1. Loading the rotor to achieve balance

To replace rotor, first make sure the motor shaft and rotor mounting hole are clean. Place the rotor on 
the motor shaft. Reinstall the rotor securing screw on the motor shaft by turning it clockwise. Hold the 
rotor with one hand and 

tighten the rotor securing screw, using the rotor wrench

6.4 Loading the Rotor

Tubes to be loaded should be filled equally by eye. The difference in the weight between the tubes 
should not exceed 0.1 gram. Tubes should always be loaded so that there is equal spacing between all 
tubes. One or two additional loaded tubes may need to be added to achieve this. Refer to Figure 1. to see 
one typical balancing scheme. 

6.5 Overloading Rotors

The maximum load of the rotor and maximum speed has been established by the manufacturer. Do 
not attempt to exceed these values. The maximum speed of the rotor has been established for liquids 
having a homogeneous density of 1.2g/ml or less. In order to centrifuge liquids with a higher density 
it is necessary to reduce the speed. 

Failure to reduce the speed may result in damage to the rotor and 

centrifuge

. The revised maximum speed can be calculated with the following formula: 

                                   1.2
                  higher density value 

Reduced speed (n

red

) =                                               × max speed (n

max

)

                   1.2
                   1.7 

(n

red

) =                × 15,000 = 12,602 rpm

Example: 

Where the density of the liquid is 1.7, the new maximum speed would be calculated as follows:

                                   1.2
                  higher density value 

Reduced speed (n

red

) =                                               × max speed (n

max

)

                   1.2
                   1.7 

(n

red

) =                × 15,000 = 12,602 rpm

If in doubt concerning maximum speeds, please contact the manufacturer for assistance. 

Содержание Prism

Страница 1: ...Prism Microcentrifuge Instruction Manual Catalog Numbers C2500 C2500 230V This manual is available in additional languages at www labnetlink com...

Страница 2: ...sories 3 6 2 Rotor Maintenance 4 6 3 Removing and Installing the Angle Rotor 4 6 4 Loading the Rotor 4 6 5 Overloading Rotors 4 7 0 Operation 5 7 1 Attaching Rotor Lid 5 7 2 Closing the Lid 5 7 3 Lid...

Страница 3: ...equipment Never centrifuge flammable explosive or corrosive materials Never centrifuge hazardous materials outside of a hood or proper containment facility Always load the rotor symmetrically Each tu...

Страница 4: ...nual 1 each Warranty card 1 each Power cord 1 each 5 0 Installation The centrifuge should be installed on a rigid even surface such as a stable laboratory bench countertop etc To guarantee sufficient...

Страница 5: ...ncluded Cat No AR110 Strip Spin Adapter Cat No C2400 SS Tubes Accepted 0 2 mL tubes or 8 x 0 2 mL strips Maximum Speed 15 000 rpm Centrifuge Radius Range 4 32 cm to 5 33 cm RCF G value Range 10 874 x...

Страница 6: ...otor wrench 6 4 Loading the Rotor Tubes to be loaded should be filled equally by eye The difference in the weight between the tubes should not exceed 0 1 gram Tubes should always be loaded so that the...

Страница 7: ...ARNING Do not attempt to open the lid of any centrifuge until the rotor has come to a complete stop In the event of a power failure or malfunction it may be necessary to open the lid manually 1 Discon...

Страница 8: ...ousing rotor chamber rotor and rotor accessories clean All parts should be wiped down periodically with a soft cloth For more thorough cleaning use a neutral cleaning agent pH between 6 and 8 applied...

Страница 9: ...and or time bAL Error Message Indicates imbalance Tubes not inserted symmetrically in rotor holes Load tubes symmetrically see Section 3 4 on Loading the rotor Sample liquid in tubes not balanced Make...

Страница 10: ...e has an automatic g force conversion program so g values are automatically calculated and can be displayed on the control panel for this centrifuging radius If adapters or smaller tubes are used the...

Страница 11: ...or contact your local support office Corning s Customer Service team will help arrange local service where available or coordinate a return authorization number and shipping instructions Products rec...

Страница 12: ...r local Corning sales office www labnetlink com For a listing of trademarks visit www corning com clstrademarks All other trademarks are the property of their respective owners 2021 Corning Incorporat...

Отзывы: