Chapter 9: Troubleshooting
Product Service 1-800-522-7658
53
Original instructions
PROBLEM
CAUSE
CORRECTIVE ACTION
Contamination outside
of fume hood.
Fume hood has
improper face
velocity
Have fume hood re-certified and
check remote blower exhaust
system. Hood should have
average face velocity of 60-100
fpm depending on application.
Remote blower has
excessive vibration.
Improper motor
mount.
Review
Chapter 2: Prerequisites
and
Chapter 3: Getting Started
.
The blower should be mounted on
vibration isolators or vibration
mounting pads to isolate vibration.
Inspect wheel for
damage and
appropriate
clearance.
Replace damaged wheel or adjust
clearance.
Check for objects in
blower wheel
.
Remove objects in blower wheel.
Improper inlet
connection.
Review
Chapter 3: Getting
Started
. The blower inlet should
be installed with a vibration
damper or flexible duct
connection.
Fume hood has
improper face velocity
.
Blower not sized
properly.
Review
Chapter 7: Modifying
Your Blower
.
Blower Sizing
Size
the blower properly with
equivalent resistance method.
Refer to Chapters 4 and 7.
Blower requires
RPM adjustment or
a manual damper is
needed in the
exhaust ductwork.
Review
Chapter 3: Getting Started
Adjust the fan speed and confirm
blower performance. See Chapters
3, 4, and 7.
Содержание Intelli-Sense CS 10
Страница 36: ...Chapter 5 Using Your Blower Product Service 1 800 522 7658 32 Original instructions...
Страница 48: ...Chapter 7 Modifying Your Blower Product Service 1 800 522 7658 44 Original instructions...
Страница 58: ...Chapter 9 Troubleshooting Product Service 1 800 522 7658 54 Original instructions...
Страница 62: ...Product Service 1 800 522 7658 58 APPENDIX B BLOWER DIMENSIONS...
Страница 63: ...Appendix B Blower Dimensions Product Service 1 800 522 7658 59 Original instructions...
Страница 64: ...Appendix B Blower Dimensions Product Service 1 800 522 7658 60 Original instructions...
Страница 70: ...Appendix D Typical Intelli Sense Blower Operating Curve Product Service 1 800 522 7658 66 Original instructions...
Страница 71: ...Product Service 1 800 522 7658 67 APPENDIX E TYPICAL LABORATORY CONTROLS WITH INDIVIDUAL INTELLI SENSE BLOWERS...
Страница 72: ...Product Service 1 800 522 7658 68 APPENDIX F TYPICAL PERCHLORIC ACID HOOD INSTALLATION DIAGRAM...