Chapter 3: Getting Started
Product Service 1-800-522-7658
13
Original instructions
For Coated Steel Blowers
10" Coated Steel Blowers include a 10" by 5-1/2" rectangular outlet. A
rectangular to round Transition Adaptor, Labconco part number 4722401, adapts
the outlet connection to accept 10-inch diameter PVC ductwork Transition
Adaptor. Labconco part number 4722400 adapts the outlet on these blowers to
accept 8-inch diameter PVC ductwork.
12" Coated Steel Blowers include a 13-1/2" by 7" rectangular outlet. Transition
Adaptor, Labconco part number 7003400, adapts the outlet of these blowers to
accept 12-inch diameter PVC ductwork.
For Fiberglass Blowers
10" Fiberglass Blowers feature a 10-3/4" ID outlet connection. Ten-inch diameter
ductwork will slip into this connection and should be held by sheet metal screws
through the housing. Silicone sealant should be used to seal any air leaks between
the duct and blower outlet connection.
12" Fiberglass Blowers feature a 12-3/4" ID outlet connection. Twelve-inch
diameter ductwork will slip into this connection and should be held by sheet metal
screws through the housing. Silicone sealant should be used to seal any air leaks
between the duct and blower outlet connection.
For PVC Blowers
10" PVC Blowers feature a 10-3/4" ID outlet connection. Ten-inch diameter
ductwork will slip into this connection and should be held by sheet metal screws
through the housing. Silicone sealant should be used to seal any air leaks between
the duct and blower outlet connection.
12" PVC Blowers feature a 12-3/4" ID outlet connection. Twelve-inch diameter
ductwork will slip into this connection and should be held by sheet metal screws
through the housing. Silicone sealant should be used to seal any air leaks between
the duct and blower outlet connection.
CS
FRP
PVC
Содержание Intelli-Sense CS 10
Страница 36: ...Chapter 5 Using Your Blower Product Service 1 800 522 7658 32 Original instructions...
Страница 48: ...Chapter 7 Modifying Your Blower Product Service 1 800 522 7658 44 Original instructions...
Страница 58: ...Chapter 9 Troubleshooting Product Service 1 800 522 7658 54 Original instructions...
Страница 62: ...Product Service 1 800 522 7658 58 APPENDIX B BLOWER DIMENSIONS...
Страница 63: ...Appendix B Blower Dimensions Product Service 1 800 522 7658 59 Original instructions...
Страница 64: ...Appendix B Blower Dimensions Product Service 1 800 522 7658 60 Original instructions...
Страница 70: ...Appendix D Typical Intelli Sense Blower Operating Curve Product Service 1 800 522 7658 66 Original instructions...
Страница 71: ...Product Service 1 800 522 7658 67 APPENDIX E TYPICAL LABORATORY CONTROLS WITH INDIVIDUAL INTELLI SENSE BLOWERS...
Страница 72: ...Product Service 1 800 522 7658 68 APPENDIX F TYPICAL PERCHLORIC ACID HOOD INSTALLATION DIAGRAM...