Chapter 7: Modifying Your Blower
Product Service 1-800-522-7658
38
Original instructions
Typical Fume Removal System
This diagram details the many components that are needed to complete a typical
fume removal system. All of these components as well as others are available
from Labconco.
*Minimum of 5 to 15 feet of duct in both locations is recommended for noise reduction.
Zero Pressure
Weathercap
Thermoplastic Duct
Intelli-Sense Coated
Steel, Fiberglass or
PVC Blower
Flexible Duct Connection
Thermoplastic Duct*
Thermoplastic Duct*
90° Elbow
Manual Duct Damper
(Highly recommended)
Protector
®
Laboratory Hood
SpillStopper™ Work Surface
Protector
®
Standard,
Acid or Solvent
Storage Cabinets
Speed
Control Box
Blower Transition Adapter
(for Coated Steel Blowers only)
3-5 duct diameter lengths
(approximately 3-5 feet) between
elbow and blower is required
Содержание Intelli-Sense CS 10
Страница 36: ...Chapter 5 Using Your Blower Product Service 1 800 522 7658 32 Original instructions...
Страница 48: ...Chapter 7 Modifying Your Blower Product Service 1 800 522 7658 44 Original instructions...
Страница 58: ...Chapter 9 Troubleshooting Product Service 1 800 522 7658 54 Original instructions...
Страница 62: ...Product Service 1 800 522 7658 58 APPENDIX B BLOWER DIMENSIONS...
Страница 63: ...Appendix B Blower Dimensions Product Service 1 800 522 7658 59 Original instructions...
Страница 64: ...Appendix B Blower Dimensions Product Service 1 800 522 7658 60 Original instructions...
Страница 70: ...Appendix D Typical Intelli Sense Blower Operating Curve Product Service 1 800 522 7658 66 Original instructions...
Страница 71: ...Product Service 1 800 522 7658 67 APPENDIX E TYPICAL LABORATORY CONTROLS WITH INDIVIDUAL INTELLI SENSE BLOWERS...
Страница 72: ...Product Service 1 800 522 7658 68 APPENDIX F TYPICAL PERCHLORIC ACID HOOD INSTALLATION DIAGRAM...