Refer to the following table if your blower fails to operate properly. If the
suggested corrective actions do not solve your problem, contact Labconco for
additional assistance.
PROBLEM
CAUSE
CORRECTIVE ACTION
Remote blower won’t
operate.
Wires not connected
at junction boxes or
switches.
Check connection of switches.
Check connection to control box
on top of fume hood, connection
of switch control wires, and
connection of shielded control
wires from roof blower. If needed,
replace control box on top of fume
hood.
Circuit breakers
tripped in building
electrical supply.
Reset circuit breakers.
Blower wiring is
disconnected.
Inspect blower wiring and 2 or 3
position switch for manual
operation. Inspect Intelli-Sense
Hood limit switch wiring for
automatic operation per Chapter 8.
Blower motor is
defective.
Check connection of ECM motor
cables or replace ECM motor.
Shaft coupling is
loose.
Tighten shaft coupling with
lockwashers between ECM motor
and blower wheel.
Содержание Intelli-Sense CS 10
Страница 36: ...Chapter 5 Using Your Blower Product Service 1 800 522 7658 32 Original instructions...
Страница 48: ...Chapter 7 Modifying Your Blower Product Service 1 800 522 7658 44 Original instructions...
Страница 58: ...Chapter 9 Troubleshooting Product Service 1 800 522 7658 54 Original instructions...
Страница 62: ...Product Service 1 800 522 7658 58 APPENDIX B BLOWER DIMENSIONS...
Страница 63: ...Appendix B Blower Dimensions Product Service 1 800 522 7658 59 Original instructions...
Страница 64: ...Appendix B Blower Dimensions Product Service 1 800 522 7658 60 Original instructions...
Страница 70: ...Appendix D Typical Intelli Sense Blower Operating Curve Product Service 1 800 522 7658 66 Original instructions...
Страница 71: ...Product Service 1 800 522 7658 67 APPENDIX E TYPICAL LABORATORY CONTROLS WITH INDIVIDUAL INTELLI SENSE BLOWERS...
Страница 72: ...Product Service 1 800 522 7658 68 APPENDIX F TYPICAL PERCHLORIC ACID HOOD INSTALLATION DIAGRAM...