Chapter 4: Performance Data & Safety Precautions
Product Service 1-800-522-7658
29
Original instructions
Speed
(RPM)
Optimum Airflow (CFM) and Static Pressure (inches of water)
1500 MAX
900 CFM @ 1.9" to 1025 CFM @ 1.8"
1400
900 CFM @ 1.8" to 1180 CFM @ 1.7"
1250
850 CFM @ 1.37" to 1325 CFM @ 0.8"
1000
675 CFM @ 0.9" to 1100 CFM @ 0.6"
750
450 CFM @ 0.45" to 825 CFM @ 0.33"
500
300 CFM @ 0.20" to 550 CFM @ 0.12"
300
200 CFM @ 0.08" to 400 CFM @ 0.10"
Practical range is 200 CFM, 0.08" to 1180 CFM, 1.7"
0
0.25
0.5
0.75
1
1.25
1.5
1.75
2
2.25
150
250
350
450
550
650
750
850
950 1050 1150 1250 1350 1450 1550
S
tati
c
P
ressure
(i
n.
w
.g.)
Airflow Rate (CFM)
Models 71834-12, -13, -22, -23
120- 240 Volts 10.0 Amps 1 HP
Blower Performance Curve with 12" PVC inlet
500
750
1000
RPM
1250 RPM
1500
Optimum Blower
Performance
Note:
Ambient Temperatures
Inside Motor Housing Greater
than 125
°F May Reduce
Maximum Blower Performance
See Note
Содержание Intelli-Sense CS 10
Страница 36: ...Chapter 5 Using Your Blower Product Service 1 800 522 7658 32 Original instructions...
Страница 48: ...Chapter 7 Modifying Your Blower Product Service 1 800 522 7658 44 Original instructions...
Страница 58: ...Chapter 9 Troubleshooting Product Service 1 800 522 7658 54 Original instructions...
Страница 62: ...Product Service 1 800 522 7658 58 APPENDIX B BLOWER DIMENSIONS...
Страница 63: ...Appendix B Blower Dimensions Product Service 1 800 522 7658 59 Original instructions...
Страница 64: ...Appendix B Blower Dimensions Product Service 1 800 522 7658 60 Original instructions...
Страница 70: ...Appendix D Typical Intelli Sense Blower Operating Curve Product Service 1 800 522 7658 66 Original instructions...
Страница 71: ...Product Service 1 800 522 7658 67 APPENDIX E TYPICAL LABORATORY CONTROLS WITH INDIVIDUAL INTELLI SENSE BLOWERS...
Страница 72: ...Product Service 1 800 522 7658 68 APPENDIX F TYPICAL PERCHLORIC ACID HOOD INSTALLATION DIAGRAM...