6. Convenient & Useful Functions
40
Menu
Submenu
Description
Outline
Effective work by skipping setups and data
manipulation on a number of different
samples is made possible, and convenient
tools are provided including sample files for
continuous measurement both in single mode
and multi-mode with sample changer, the
auto-mode of barcode reader, concentration
scanning to a create table and temperature
scanning to create compensation formula.
6-2-1
Sample mode
Select a function out of them available in a
sample.
6-2-2
Single mode
The initial mode is Single mode (default).
The lowest digit of sample number is
incremented each time of measurement.
6-2-3
Sample file
Easy to set up measure condition by selecting
registered sample file. You can select,
register, copy or delete sample file.
6-2-4
Auto mode
Auto mode is effective when selected for
sample mode. Useful with barcode reader.
(See 6-9-4 Barcode reader)
Barcode reader can call for method name and
file, user name, etc.
6-2-5
Concentration scan
This mode creates working curve for
concentration
from
density
or
SG.
Concentration table (relation of concentration
to density) is created and saved easily by
measuring a few samples of which
concentration is known. The concentration of
unknown sample can also be calculated from
density in the table.
6-2
Sample
6-2-6
Temperature scan
Thermal character of sample (liquid
temperature vs. density) is obtained by this
mode. Temperature compensation formula
can be easily created from a few thermal
points and shown as a result of
measurement. Calculation of density at
temperatures
different
from
actual
measurement can be made on this table.