5. Measurement
32
3
Sampling
z
Adjust sample/drain lever to “Sample.”
z
Adjust sampling speed controller to
slowest (scale 0).
z
Press “PUMP” to start sampling.
z
Slowly turn the controller until you reach
optimal speed for sampling.
z
Check through cell window to make sure
there is no air bubble in the cell.
z
Press “PUMP” again to stop sampling.
Hint
Be careful when sampling speed is increased abruptly (e.g.
scale 0
Æ
5), air bubbles may be formed.
Hint
For a series of measurements of the same sample, continue
sampling at the same speed. For a different type of sample,
start sampling by adjusting speed first.
4
Start measurement
z
Press “START”.
A message of “Measuring” will
appear.
5
Measurement ends
z
Once oscillation frequency
becomes stable, measurement
comes to an end and result
(density of sample) will be
shown.
Result will be printed out
when printer is connected.
Check air bubbles.