LCD indication/ Battery level indicator
KEW 2062/2062BT
- 13 -
KEW 2062/2062BT
Screw
Battery compartment cover
Size AA Alkaline battery: LR6
CAUTION
●
Brand and type of the batteries should be harmonized.
●
Never mix new and old batteries.
●
Install batteries in correct polarity as marked inside the battery compartment area.
LCD indication/ Battery level indicator
Ba
tte
ry
le
ve
l
Status Details
Battery level is full.
Indicator varies depending on the battery level.
Battery level is low. Replace the batteries with new
ones.
Blink
Battery level is extremely low, and the instrument
doesn’t work normally. Stop using the instrument and
replace the batteries with new ones immediately.
The instrument continues measurement even in this
state; however, Bluetooth will be disabled.
How to install batteries:
Follow the procedures below and insert batteries.
1 Disconnect all cables and set the function switch to OFF position.
2 Loosen one battery compartment cover-fixing screw and remove the cover.
3 Take out all the batteries.
4 Insert two new batteries, size AA alkaline: LR6, observing correct polarity.
5 Install the cover, and then secure it with the screw.
KEW 2062/2062BT
4.3 Test leads connection (to KEW 2062 / 2062BT)
KEW 2062/2062BT
- 14 -
4.3 Test leads connection (to KEW 2062 / 2062BT)
The followings should be checked before the connection.
DANGER
●
Use only the test leads supplied with this instrument.
●
Connect the cables that are required for the desired measurement only.
●
First, connect the plug of the test lead to the instrument. Only then connect to the
measurement line.
●
Never disconnect the test lead from the voltage input terminal of the instrument during a
measurement (while the instrument is energized).
WARNING
●
Never attempt to make measurement if any abnormal conditions are noted, such as a crack
or exposed metal parts.
CAUTION
● Confirm that the instrument is powered off, and
then connect the test leads.
●
Connect to the instrument first, firmly into the corresponding terminal.
Connect the test leads per the following procedures.
1 Ensure that KEW 2062 / 2062BT is turned off.
2 Connect the test lead to AC voltage input terminal
on the instrument.*
* The number of test leads to be connected
depends on the wiring configuration.
4.4 Connection to the measured object
The following should be checked before the connection.
DANGER
●
This instrument is rated to 300 V AC for CAT IV, 600 V AC for CAT III, and 1000V AC for
CAT II. With attention to the measurement category to which the object under test
belongs, do not make measurements on a circuit in which voltage exceeds these values.
● Use only the test leads des
igned for this instrument.
●
Always connect the test leads to the instrument first.
●
When the instrument and the test lead are combined and used together, whichever
lower category either of them belongs to will be applied. Pay attention to the rating of
the instrument and test lead used together.
●
Connect the cables that are required for the desired measurement only.
●
Current sensor shall be connected to the secondary side of the circuit breaker since the
primary side has dangerous large current capacity.