Chapter 3 Basic operation
KEW 2062/2062BT
- 9 -
KEW 2062/2062BT
Chapter 3 Basic operation
3.1 Function switch
Function
Description
SETUP
Settings
Changes and confirms the settings for wiring, VT/ CT ratio,
buzzer on/ off, backlight on/ off, nominal frequency 50/ 60Hz. To
restore all the settings to the default conditions, perform system
reset.
Phase
detection
Tests and shows the phase rotation sequence, and missing
phase if there is any.
Harmonics
Shows voltage/ current (of the 1st fundamental waveform up to
30th harmonics) RMS value, content rate, and distortion factor
[THD-R/THD-F].
W
Power
Shows: active/ reactive/ apparent power, power factor,
voltage-current phase differences, and voltage/ current (RMS)
value.
~
V
AC voltage Shows AC voltage RMS value, peak value, and frequency.
~
A
AC current
Shows AC current RMS value, peak value, and frequency.
3.2 Buttons and switches
Function Buttons and
switches
Details
---
Data hold button
Hold down the Data hold button until the LCD shows “
”
symbol: then the currently displayed value is held. While this
function is activated, readings don’t change even the input
value varies. To exit from the hold mode, press the data hold
button again or switch the measurement functions: the “
”
symbol disappears.
Backlight button
[
◄
] A long press turns on/ off the backlight.
SETUP
Item switching
button
[
◄►
] Toggles displayed items and changes the setting values.
Mode button
Selects setting items and confirms the entered values.
Harmonics
Item switching
button
[
◄►
]
A short press toggles display: <-> THD-F <-> THD-R <->
1st fundamental wave to 30th harmonics.
[
►
] A long press switches voltage and current RMS values.
Mode button
A short press toggles display: <-> Inst value <-> MAX <-> MIN
<-> AVG.
A long press resets measurements of MAX, MIN, and AVG
values and resume a measurement.
KEW 2062/2062BT
3.2 Buttons and switches
KEW 2062/2062BT
- 10 -
Function
Buttons and
switches
Details
Power
1P2W
1P3W
Item switching
button
[
◄►
]
A short press toggles display: <-> active power, power factor
<-> active power, voltage-current phase differences <->
active and apparent power <-> active and reactive power
<-> current and voltage RMS.
Mode button
A short press toggles display: <-> Inst value <-> MAX <->
MIN <-> AVG.
A long press resets measurements of MAX, MIN, and AVG
values and resume a measurement.
Power
3P3W
3P4W
Balance
Item switching
button
[
◄►
]
A short press toggles display: <-> active power, power factor
<-> active and apparent power <-> active and reactive
power <-> current and voltage RMS.
Mode button
A short press toggles display: <-> Inst value <-> MAX <->
MIN <-> AVG. A long press resets measurements of MAX,
MIN, and AVG values and resume a measurement.
Power
3P3W
Unbalance
Item
switching
button
[
►
] A short press during a measurement:
Switches the phase to be measured from R(L1) to T(L3).
[
◄►
]
A short press while displaying the measured result:
Toggles displays: <-> Three-phase active power <->
R(L1)-phase active power <-> T(L2)-phase active power.
Mode button
A short press during a measurement: Switches between
active power and voltage and current values (RMS).
A long press while displaying the measured result:
Clears the displayed values and resume a measurement.
Power
3P4W
Unbalance
Item switching
button
[
►
]
A short press during a measurement:
Switches the phase to be measured: R(L1) -> S(L2) -> T(L3).
A short press while displaying the measured result:
Toggles displays: <-> active power, power factor <-> active
and apparent power <-> active and reactive power.
Mode button
A short press during a measurement: Switches between
active power and voltage and current values (RMS).
A long press while displaying the measured result:
Clears the displayed values and resume a measurement.
~
V
~
A
Mode button
A short press toggles display: <-> Inst value <-> MAX <->
MIN <-> AVG <-> |PEAK| (peak value*).
A long press resets measurements of MAX, MIN, AVG, and
|PEAK| values and resume a measurement.
* |PEAK|: Shows the instantaneous peak value in absolute
value.