How to Enter Calibration Data
3-6
965X Operation Manual
How
to Enter Calibration Data
Adjustment menus are located in the setup screens. In these menus, you can add negative or posi
-
tive adjustments to move a servo to desired positions to compensate (calibrate) for any deviation’s
from the expected cut. The calibration can then be saved.
Note:
Maximum offset is 0.50 inches. If more than 0.50 inches of adjustment is
needed, there is a high probability that another issue may be causing a problem.
If issues can not be resolved, contact the
Kval Service Center
(1-800-553-5825).
About The Calibration Box
Below are general instructions to perform a calibration.
1.
Check for deviations between the expected machining and the actual machining.
Determine the offset.
2.
Enter a positive or negative offset in the
Offset
box to adjust the servos in the X, Y,
and Z directions.
Note:
Select the
Combine Button
to keep track of the offsets during the process.
3.
After every calibration change, select the
Save
button to store the calibration. Run a
test jamb again to verify the offset correction was successful.
4.
Keep running a test jamb and adjusting the offset until satisfied with the result.
Total: Is the running
total of the Base and
the Offset
Combine: Merges the
offset number to the
base.
Base: Is the result of
the Combine selec
-
tion. (The merge of
the Offset and the
Base)
Offset: Enter the
desired offset to bring
into calibration.
Save Button: Select to store the calibration.
Содержание 965X
Страница 29: ...How to Download the Service Application 1 21 965X Operation Manual Page Intentionally Left Blank ...
Страница 103: ......