Remark :

A. If    the  current  value  is  unknown  before 

measurement, select the highest range.

B. When  only  “1”  occur  that  indicates  the  tested 

current  has  exceeded  the  selected  range.   
You should select a higher range.

C. When doing mA jack input, overload may melt 

the fuse inside.   Now replace the Fuse.   The 
Fuse specifications is 0.2A / 250V.

D. When  doing  20A  jack  input,  time  for  inputing 

the highest current 20A should not exceed 15 
seconds.

4.  AC Current Measurement :

COM

V

W

Hz

20A

mA

NPN

B

B

E

E

E

PNP

E

C

C

Cx

KM 6050

POWER

LIGHT

HOLD

Auto 
Power off

200

2K 20K 200K

2M

20M

200M

200m

2

V

20

200

1000

1000

200

20

V

2

2n

20n

200n

2

m

200

m

2m

20m

200m 

20

20

200m 

20m

2m

A

hFE

20kHz

A

W

F

(1) Remove the test lead and set the function / range 

switch on ACA range position.  Refer to DCA 
measurement (1).

(2) Insert the black test lead into COM jack and the 

red test lead into the exposed test lead jack (mA 
jack or 20A jack). Connect the test lead to the 
tested circuit in series.

Supply

Load

Remark :

A. Refer to DCV measurement remarks

     A, B,C, D.

B. “     ” indicate not to measure the voltage over 

1000V.  Although it is possible to display the 
reading, it will probably damage the meter.

!

3.  DC Current Measurement :

(1) Remove the test lead and set the function / 

range switch on DCA range.

(2) Insert the black test lead into COM jack and 

the red one into the exposed test lead jack 
(mA jack or 20A jack).  Connect the test lead 
to the tested circuit in series.  The meter will 
indicate the polarity of the red test lead at the 
same time as the meter indicate the current 
reading.

COM

V

W

Hz

20A

mA

NPN

B

B

E

E

E

PNP

E

C

C

Cx

KM 6050

POWER

LIGHT

HOLD

Auto 
Power off

200

2K 20K 200K

2M

20M

200M

200m

2

V

20

200

1000

1000

200

20

V

2

2n

20n

200n

2

m

200

m

2m

20m

200m 

20

20

200m 

20m

2m

A

hFE

20kHz

A

W

F

12

Supply

Load

Содержание KM 6050

Страница 1: ...alog Digital Insulation Resistance Testers upto 10 KV Digital Sound Level Meter Sound Level Calibrator Digital contact Non contact Type Tachometer Digital Non contact infrared Thermometer Thermo Hygro...

Страница 2: ...we damage them or burn them out If you incorrectly...

Страница 3: ...cient protection for the Function Range switch You can select different models to measure DCV DCA ACV ACA Resistance Capacitance Frequency Temperature Diode hFE and Continuity This type can be applied...

Страница 4: ...could result in serious injury or even death to the user Terms in this manual Low power consumption CMOS double integration A D transform integrated circuit Auto zero Calibration Auto polarity displa...

Страница 5: ...Frequency Range 40Hz 400Hz Overload protection 250V for 200mV DC or AC peak value 1000V for other ranges Display Average value Sine RMS 4 DC CURRENT Range Accuracy Resolution 2 mA 20 mA 200 mA 20 A 10...

Страница 6: ...ge measurement Approx 40mV 2 5 10 0 5 10 0 5 5 0 2 1 2 m F 200 m F 6 FREQUENCY Range Accuracy Resolution 20 kHz 1Hz Voltage Protection 250V DC or AC peak value Input sensitivity 150 mV RMS 2 5 TRIODE...

Страница 7: ...d Danger High Voltage 8 9 COM VW Hz 20A mA NPN B B E E E PNP E C C Cx 6 MULTIMETER STRUCTURE 4 5 6 1 7 2 3 8 1 LCD display 2 Function range selection 3 Test lead jack 4 Power switch 5 Triode measureme...

Страница 8: ...1000 200 20 V 2 2n 20n 200n 2m 200m 2m 20m 200m 20 20 200m 20m 2m A hFE 20kHz A W F 10 2 AC Voltage Measurement 1 Set the function range switch on ACV range 2 Insert the black test lead into COM jack...

Страница 9: ...2 Insert the black test lead into COM jack and the red test lead into the exposed test lead jack mA jack or 20A jack Connect the test lead to the tested circuit in series Supply Load Remark A Refer t...

Страница 10: ...he positive polarity of the measurement circuit Connect the test lead to the two ends of the tested resistance 15 6 Capacitance measurement 1 Set the function range switch on the range position needed...

Страница 11: ...surement unit 1pF 10 m F 1nF 10 m F D Don t insert the capacitors with high voltage into the test jack direct 17 B If the tested signal is too strong you should use the external attenuation so as not...

Страница 12: ...NPN B B E E E PNP E C C Cx KM 6050 POWER LIGHT HOLD Auto Power off 200 2K 20K 200K 2M 20M 200M 200m 2 V 20 200 1000 1000 200 20 V 2 2n 20n 200n 2m 200m 2m 20m 200m 20 20 200m 20m 2m A hFE 20kHz A W F...

Страница 13: ...tion 2 Insert the black test lead into COM socket and the red one into VW Hz jack and connect the test lead to the two ends of the circuit to be tested Remark You must cut off the power before you che...

Страница 14: ...test leads from the input terminals Remove the back cover screws and open the back cover Replace the fuse fuses with new fuse of the same type and rating 0 2A 250V 6 Remove the battery and put it in a...

Страница 15: ...ntees that the product has been inspected and tested in accordance with the published specifications The instrument has been calibrated by using equipment which has already been calibrated to standard...

Страница 16: ...ed dealer shall extend this warranty on new and unused products to end user customers only but have no authority to extend a greater or different warranty on behalf of KUSAM MECO KUSAM MECO s warranty...

Отзывы: