4. When the test leads are plugged into the current input
terminals, do not connect the test leads to any circuit in
parallel or it will damages the fuses and the instrument.
5. After the completion of all the measurement operation,
you should first turn off the power & then disconnect the
test leads & the measured circuit connection, Especially
to high current measurement.
6. It is forbidden to input more than 36V DC, 25V AC
Voltage between the current jack & the “COM” jack.
3000
AC
mA
SELECT
HOLD
MAX/MIN
REL
RANGE
FUSED
FUSED
CAT III 600V
10secMAX
10A
MAX
600mA
750V DC
1000V DC
10A
COM
V
W
HzmA
m
A
out
AUTO POWER OFF
TRUE RMS
OFF
V
V
~
NCV
W
Hz
%
A
mA
~
~
~
m
A
out
!
SELECT
mA
~
The Maximum measured pressure drop :
600mV
Overload Protection :
60mA / 600mA : 600mA/250V Speed Glass Fuse.
AC CURRENT(AC mA)
60 mA
600 mA
10
m
A
100
m
A
±(1.2%rdg + 10dgts)
Range
Accuracy
Resolution
The specific operation is
as follows: The diagram
on the right is the DC mA
schematic diagram.
1. Insert the black test
lead into the “COM”
socket. Insert the red
test lead into the
“mA / mA” jack.
(max. 600mA), “SELECT”
key to switch between
AC & DC current.
10
2.Turn the range switch to the corresponding DC/AC mA
gear, trigger “ SELECT” to select AC m A and then put
the meter into the circuit under test. The measured
current value and the current polarity of the red test
point will also be displayed at the same time. on the
screen.
NOTE :
1.The instrument series connection to the circuit
pending
to test before the circuit should be the first power off.
2. If there is no concept of the measured current range
in advance, the range switch should be transferred to
the highest range, & then according to the display value
to the corresponding file; such as the screen display
“0L”
that has exceeded the range, needed to turn the range
switch to the appropriate gear.
3. The maximum input current is 600mA (depending on
the location of the red test pen inserted), excessive
current will damage the mA file fuse.
4.When the test leads are plugged into the current input
terminals, do not connect the test leads to any circuit in
parallel or it will damages the fuses and the instrument.
5.After the completion of all the measurement operation,
you should first turn off the power & then disconnect the
test leads & the measured circuit connection, Especially
to high current measurement.
6. It is for bidden to input more than 36V DC, 25V AC
Voltage between the current jack & the “COM” jack.