07
NOTE :
1.The instrument series connection to the circuit pending to test
before the circuit should be the first power off.
2.If there is no concept of the measured current range in advance,
the range switch should be transferred t o the highest range, &
then according to the display value to the corresponding file; such
as the screen display “0L” that has exceeded the range, needed
to turn the range switch to the appropriate gear.
3.The maximum input current is 600mA (depending on the location
of the red test pen inserted), excessive current will damage the
mA file fuse.
4.When the test leads are plugged into the current input terminals,
do not connect the test leads to any circuit in parallel or it will
damages the fuses and the instrument.
5. After the completion of all the measurement operation, you
should first turn off the power & then disconnect the test leads &
the measured circuit connection, Especially to high current
measurement.
6. It is for bidden to input more than 36V DC, 25V AC Voltage
between the current jack & the “COM” jack.
The Maximum measured pressure drop :
600mV
Overload Protection :
600
m
A / 6000
m
A : 600mA/250V Speed Glass Fuse.
DC CURRENT (
m
A)
600
m
A
6000
m
A
0.1
m
A
1
m
A
±(1.2%rdg + 10dgts)
Range
Accuracy
Resolution
08
The specific operation
is as follows:
1.Insert the black test
lead into the “COM”
socket. Insert the
red test lead into the
“mA / uA” jack.
(max. 600mA),
“SELECT” key to
switch between AC &
DC current.
2. Turn the range switch to the corresponding DC / AC
uA position, then insert the instrument into the circuit
pending to test. The current value of the measured
current & the polarity of the red test point will also be
displayed on the screen at the same time.
NOTE :
1. The instrument series connection to the circuit pend-
ing to test before the circuit should be the first power off.
2. If there is no concept of the measured current range
in advance, the range switch should be transferred to
the highest range, & then according to the display value
to the corresponding file; such as the screen display
“0L” that has exceeded the range, needed to turn the
range switch to the appropriate gear.
3.The maximum input current is 6000uA (depending on
the location of the red test pen inserted), excessive
current will damage the mA file fuse.
3.000
DC
A
SELECT
HOLD
MAX/MIN
REL
RANGE
FUSED
FUSED
CAT III 600V
10secMAX
10A
MAX
600mA
750V DC
1000V DC
10A
COM
V
W
HzmA
m
A
out
AUTO POWER OFF
TRUE RMS
OFF
V
V
~
NCV
W
Hz
%
A
mA
~
~
~
m
A
out
!
SELECT
m
A
~
m