OPERATION
6
41
WATERFLUX 3070
www.krohne.com
06/2016 - 4004999701 - MA WATERFLUX 3070 - R01 en
6.1.3 Status information on the display
Warning and error messages
Display symbols
Description
Menu setting
Battery status
Always available
AMR mode on
Available if menu number 2 is set to 1
Automatic self-test
Automatically
Please refer to
Automatic self-test
on page 62.
Test mode on
To activate
Please refer to
Test mode
on page 62.
Display
Description
Actions
At the current
consumption rate, the
battery will be empty
within 1 year.
Plan battery replacement. Note: this sign can
also appear for a short period if the battery
consumption is temporarily very high.
Battery nearly empty
Replace the battery
Battery empty:
voltage too low
(Flashing)
Warning
•
Check all connections
•
If the exclamation mark does not disappear,
contact the service department
(X = 1...127)
Impaired software
integrity
The meter has stored the last counter values
and went to sleep mode.
Contact the service department.
Empty pipe
Available if menu number 83 is set to 1.
When the
pressure/temperature
is above or below
max/min of the
specified threshold
values, the display will
show the applicable
max/min calibrated
value, blinking
ON/OFF
Check system and make sure not to exceed P/T
range.
( -0.5...16 bar / -5...+70
°
C)
When
pressure/temperature
is not read out
correctly ( or not read
out yet from the P&T
sensor) the display
will show some
dashes to indicate
there is no valid data
available
•
Check all connections
•
If the exclamation mark does not disappear,
contact the service department