OPERATION
6
61
WATERFLUX 3070
www.krohne.com
06/2016 - 4004999701 - MA WATERFLUX 3070 - R01 en
6.3.13 Modbus settings
CAUTION!
Changing the baudrate affects the battery lifetime.
For more details refer to the Modbus supplementary manual.
No.
Function
Options
Description
Modbus RS485
Modbus RS485
Modbus RS485
Modbus RS485
C0
Modbus slave address
001
Value to be programmed (1 to 247)
C1
Baud rate
Baud rate choice from table :
1.2
1200 baud
2.4
2400 baud
3.6
3600 baud
4.8
4800 baud
9.6
9600 baud (default)
19.2
19200 baud
38.4
38400 baud
57.6
57600 baud
115.2
115200 baud
C2
Parity rate
0
Even (default)
1
Odd
2
No parity
C3
Register format
0
Little endian
1
Big endian (default)
2
Mix endian, small words
3
Mix endian, large words
C4
Transmission delay
0 = 0ms
Transmission delay can be set by choice from
table
5 = 5ms
10 = 10ms
20 = 20ms
30 = 30ms
40 = 40ms
50 = 50ms
(default)
C5
Stop bits
1
1 Stop bit (default)
2
2 Stop bits
C6
User application name
00000
To be set by programmer
C7
RS485 bus termination
0
No RS485 termination (default)
1
RS485 termination
2 x 120
Ω
C8
Reception interval
0 (default)
Set value for range 0-3600 sec.