OPERATION
6
53
WATERFLUX 3070
www.krohne.com
06/2016 - 4004999701 - MA WATERFLUX 3070 - R01 en
Pulse output A and B give the same number of pulses, but pulse output B is shifted half a pulse
width in time. Pulse output B always runs behind pulse output A.
In case of a forward flow, pulse output A and B give the forward volume passing through the
meter.
In case of a reverse flow, pulse output A and B only start to give pulses again after the sum
counter has reached the previous maximum net volume.
INFORMATION!
In case of net forward volume the sum counter on the display can temporarily be different from
the net volume provided by pulse output A and B.
No.
Function
Options
Description
20
Pulse output A
0 = Off (default)
Pulse output A
1 = On
21
Pulse output B
0 = Off (default)
Pulse output B
1 = On
22
Pulse output A and B:
phase shift
0 = 90
°
offset (default)
Sets the relation between pulse
output A and B. Menu no. 20 and 21
must be set to 1 = on.
1 = Forward - reverse
2 = Net forward volume
23
Pulse output A and B:
pulse width
5 = 5 ms (default)
The pulse width is always the same
for pulse output A and B.
A small pulse width means a lower
energy consumption of the battery.
10 = 10 ms
20 = 20 ms
50 = 50 ms
100 = 100 ms
200 = 200 ms
24
Pulse output A and B:
pulse value
00100 = litre
Unit and format are automatically
set based on the selection made in
menu number 10.
00.100 = m
3
/pulse
0010.0 = gallon
0010.0= imp. gallon
001.00 = ft
3
.00100 = a-inch
.00100 = a-ft
0.1000 = m
3
/pulse
When menu item A2 is set on 1
25
Pulse output A and B:
pulse generation
99 = Automatically set (default)
Pulse output reacts as set in menu
number 24
1
*******1 (A pulse is generated if the
least significant digit changes)
2
******2*
3
*****3**
4
****4***
5
***5****
6
**6*****
7
*7******
8
8******* (A pulse generated if the
most significant digit changes)