3
ELECTRICAL CONNECTIONS
30
POWERFLEX 2200 C/F/S/D
www.krohne.com
07/2016 - 4005318901 - QS POWERFLEX 2200 R01 en
3.1.2 Remote version
For more electrical installation data, refer to
Compact version
on page 28.
Terminals for electrical installation
Figure 3-5: Terminals for electrical installation
1
Grounding terminal in the housing (if the electrical cable is shielded)
2
Current output -
3
Current
4
Location of the external grounding terminal (on the wall support)
INFORMATION!
Electrical power to the output terminal energizes the device. The output terminal is also used for
HART
®
communication.
CAUTION!
•
Use the applicable electrical cables with the cable glands.
•
Make sure that the current is not more than 5 A or that there is 5 A-rated fuse in the
electrical circuit that energizes the device.
Connections between the remote converter and the probe housing
Figure 3-6: Connections between the remote converter and the probe housing
1
Remote converter
2
Probe housing
3
Power supply: voltage in -
4
Power supply: voltage in +
5
Signal cable B
6
Signal cable A
7
Shielding wire (attached to Faston connectors in the housings of the remote converter and the probe housing)