SERVICE
6
73
OPTISYS SLM 2100
www.krohne.com
01/2017 - 4002737304 - MA OPTISYS SLM 2100 R04 en
Visual check and cleaning of lens
Check the lens for any build-up or film. If necessary, clean it with a suitable agent like water or
isopropyl alcohol and a soft rag or a soft brush.
Check of screws and nuts
Check if all screws of the reeling unit as well as on the mounting frame are properly tight and
secured.
Sensor check in water
Check the sensor by immersion in water; this should give a low measuring value or even a "0" on
the converter display. By placing something solid (e.g. your finger) in the gap between the LED
and the sensor you should get a measuring value of 3100...3200 (depending on the calibration).
Visual check of cable and drum
Check the cable, it must not have any damages or wear and has to be spooled correctly on the
drum. Always replace the cable if it shows any damage or severe wear of the insulation!
If the cable is not correctly spooled up or shows a large amount of dirt, lower the sensor
manually to the maximum position at first. During back hauling monitor the correct guidance of
the cable and that it spools up properly, clean it with a soft cloth if necessary.
Cleaning of the optical interface
•
Open the cable compartment door.
•
Pull the pick-up arm back and hold it.
•
Clean the optical interface with damp soft cloth with isopropyl alcohol.
•
Close the pick-up arm carefully and make sure that the connection is closed.
•
Close the cable compartment door.
WARNING!
Open and close the pick-up arm carefully to prevent the hazard of squashing and bruising.
Follow the steps for the maintenance of the optical interface.
Содержание OPTISYS SLM 2100
Страница 5: ...CONTENTS 5 www krohne com 01 2017 4002737304 MA OPTISYS SLM 2100 R04 en OPTISYS SLM 2100 8 Notes 85 ...
Страница 85: ...NOTES 8 85 OPTISYS SLM 2100 www krohne com 01 2017 4002737304 MA OPTISYS SLM 2100 R04 en ...
Страница 86: ...8 NOTES 86 OPTISYS SLM 2100 www krohne com 01 2017 4002737304 MA OPTISYS SLM 2100 R04 en ...
Страница 87: ...NOTES 8 87 OPTISYS SLM 2100 www krohne com 01 2017 4002737304 MA OPTISYS SLM 2100 R04 en ...