4
ELECTRICAL CONNECTIONS
38
OPTISYS SLM 2100
www.krohne.com
01/2017 - 4002737304 - MA OPTISYS SLM 2100 R04 en
•
Open the cable drum and electronic compartment door.
•
Push the prepared cables through the cable feedthrough and route them to the electronic
compartment (for more information refer to
Overview of cable connections
on page 30).
•
Connect the positive and negative lead according to the connection diagrams above.
•
Connect the shield to one side only (e.g. to the protective earth on the mainboard next to the
relay) side.
•
Tighten the screw connection of the cable feedthrough securely.
•
Seal all cable feedthrough that are not needed with a plug.
•
Close both compartment doors
4.9.3 Electrical connection of control inputs
INFORMATION!
All cables must have a test voltage of min. 2 kV and an appropriate outer insulation (additional to
the insulation of the individual wire) . The outer insulation should be removed 30...50 mm /
1.18...1.97 inch. The minimum allowed cross section of the wires is 0.5 mm
2
/ 20 AWG and the
maximum is 1.5 mm
2
/ 16 AWG.
INFORMATION!
In order to assure proper sealing of the cable feedthrough only cables with a diameter between
6...12 mm / 0.24...0.47 inch should be used.
DANGER!
All work on the electrical connections may only be carried out with the power disconnected.
CAUTION!
To avoid damage or destruction of the device always note the properties of the control input; for
further information refer to Technical data on page 78
.
INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.
INFORMATION!
All cables must have a test voltage of min. 2 kV and an appropriate outer insulation (additional to
the insulation of the individual wire) . The outer insulation should be removed 30...50 mm /
1.18...1.97 inch. The minimum allowed cross section of the wires is 0.5 mm
2
/ 20 AWG and the
maximum is 1.5 mm
2
/ 16 AWG.
Содержание OPTISYS SLM 2100
Страница 5: ...CONTENTS 5 www krohne com 01 2017 4002737304 MA OPTISYS SLM 2100 R04 en OPTISYS SLM 2100 8 Notes 85 ...
Страница 85: ...NOTES 8 85 OPTISYS SLM 2100 www krohne com 01 2017 4002737304 MA OPTISYS SLM 2100 R04 en ...
Страница 86: ...8 NOTES 86 OPTISYS SLM 2100 www krohne com 01 2017 4002737304 MA OPTISYS SLM 2100 R04 en ...
Страница 87: ...NOTES 8 87 OPTISYS SLM 2100 www krohne com 01 2017 4002737304 MA OPTISYS SLM 2100 R04 en ...