5.7.4 Assignment of current outputs
The function of outputs is different regarding which measuring function has been selected. It is
changing automatically if the measuring function has been changed as outlined in the following
table.
5.7.5 Measuring mode
The level measurement can be set in two different ways:
1. Height: The height of the sludge blanket/fluff measured from the bottom of the tank is dis-
played. Zero (resp. 4 mA) is at the bottom of the tank.
2. Depth: The depth of the sludge blanket/fluff measured from the liquid surface is displayed. Zero
(resp. 4 mA) is at the liquid surface or at the home position.
The height / depth ratio of the sensor is calculated in the following way:
Height /depth = maximum immersion depth (- blind zone) - start position.
Selection of measuring functions in menu C1.2
Measuring function
Level
Profile
Zone tracking
Current output A
Height / depth of
concentration (fluff)
1
Height / depth of the
sensor position
1
Height / depth of the
sensor position
1
Current output B
Height / depth of
concentration (sludge
blanket)
1
Measured concentration
2
Measured concentration
2
1
Range is calculated automatically according to the parameters filled in the extended setup; range of analog output is
shown in the I/O setup of the device.
2
Concentration range of the output must be defined in the I/O setup of the device.
Press >>>> for more than 2.5 seconds, then release the button.
Press
↓
or
↑
until the main menu setup
setup
setup
setup is highlighted.
Press >>>> to enter the chosen menu.
Press
↓
or
↑
to select measuring mode
measuring mode
measuring mode
measuring mode.
Press >>>> to enter the chosen menu.
Press
↓
or
↑
to select height
height
height
height, or depth
depth
depth
depth.
Press
^
^
^
^
to confirm.
Press
^
^
^
^
several time until you reach the measuring mode again. Choose yes
yes
yes
yes to safe and confirm your
selection.
INFORMATION!
If the blind zone should be included in this calculation ensure that "yes" is selected in menu C4.4.
Содержание OPTISYS SLM 2100
Страница 5: ...CONTENTS 5 www krohne com 01 2017 4002737304 MA OPTISYS SLM 2100 R04 en OPTISYS SLM 2100 8 Notes 85 ...
Страница 85: ...NOTES 8 85 OPTISYS SLM 2100 www krohne com 01 2017 4002737304 MA OPTISYS SLM 2100 R04 en ...
Страница 86: ...8 NOTES 86 OPTISYS SLM 2100 www krohne com 01 2017 4002737304 MA OPTISYS SLM 2100 R04 en ...
Страница 87: ...NOTES 8 87 OPTISYS SLM 2100 www krohne com 01 2017 4002737304 MA OPTISYS SLM 2100 R04 en ...